UV-C STERILON FLOW SAFETY INSTRUCTIONS
FLOW VERSION
UVC flow luminaires use ultraviolet light, destroy the DNA or RNA of all
microorganisms that will be exposed and remove viruses, bacteria and fungi from the
air. Covered UVC light does not reach the outside of the lamp. Thanks to stimulated
circulation, the air passes through luminaire interior, where it is exposed to the light
and cleaned, and then pushed outside. UV-C STERILON FLOW luminaires may work in
the presence of people.
Exposure method:
1. Place the lamp in a disinfected room, if possible in a central place, to increase the
e ectiveness of disinfection. Remember that the disinfection e ciency in a given
place decreases as the distance from the lamp increases.
2. Do not cover the luminaire fan, do not place any objects on the luminaire. In order not
to disturb circulation, leave an empty space and do not put any objects within 1m of the
luminaire fan.
3. There may be people in the room. They can move freely and do not need protective
clothing.
4. Do not open the luminaire covers during its operation.
Secure the luminaire against uncontrolled opening by other people, especially pay
attention to children.
5. Do not open the window while the luminaire is operating,
this will reduce the e ectiveness of disinfection.
6. We recommend ventilating the room every 4 operating hours of the device.
7. We recommend using the product in rooms with
unobstructed ventilation.
8. Do not use in rooms smaller than 6 m
9. The smell emitted by the UV-C STERILON FLOW is a technical fragrance. It can last
for approximately 140 hours of device use. This smell does not pose a threat to the
users of the device.
10. A small amount of visible blue light – which is a reflected light, does not contain UV-
C radiation and is safe for humans.
Disinfection e ciency
Depending on the lamp power, the flow is:
3
For the power 72 W – 160 m /h
ź
3
For the power 144 W – 200 m /h
ź
The maximum volume (cubature) of the disinfected room:
for luminaires with a power of 72 W - 75 m
ź
for luminaires with a power of 144 W - 120 m
ź
Maintenance and filter replacement
UVC - Sterilon Flow luminaires are standard equipped with PPI30 foam filters. This
filter should be cleaned under running water or replaced after every 80 hours of
luminaire operation. The filter should be removed from the push-in (click-on) housing
on the fan, replaced and the housing pressed back on. The Sterilon UV-C lamp is
designed for mounting a HEPA filter. The HEPA filter can be mounted additionally or
interchangeably with a regular filter.
DIRECT DISINFECTION
Operational safety
The UV-C light emitted by the lamp is
other living organisms, damages eyesight, skin and cell DNA
UV-C light degrades the DNA and RNA code in the cells of living organisms. After
no people or animals should be within the
starting the luminaire
range of direct UVC rays.
Damage to the luminaire will not a ect your health while maintaining
precautions.
If the fluorescent lamps break,
minutes
and remove the contaminated parts, preferably with gloves. Place them in
a sealed plastic bag and take to a local waste disposal facility for recycling. Do not use a
vacuum cleaner to remove damaged parts.
6
UV-C STERILON FLOW 144W - INSTALLATION INSTRUCTION
2
3
3
highly harmful to humans and
ventilate the room for 30
In particular, you should:
Never look directly at the turned on UV-C STERILON FLOW luminaire or stay
ź
within the reach of UV-C rays.
Secure the room with the turned on luminaire against
ź
other people's access.
When carrying out activities related to the operation of the UV-C STERILON FLOW
ź
lamp, special protective glasses and gloves as well as wearing long-sleeved
clothing are recommended (only with direct disinfection).
Disinfection e ciency
Disinfection e ciency is presented by CIE and other research units based on the dose of
energy needed to remove 90% of microorganisms from the surface. For most of the
known bacteria and viruses the required dose does not exceed 8 mJ / cm . This means
that the lamp emitting 1 W of UV-C radiation at a distance of 1 metre from the given
surface needs 80 seconds to disinfect it in 90%.
Lamps from the Sterilon series emit light at a distance of 1 metre from the source:
2
2,4 W/m - UV-C radiation (UV-C STERILON FLOW 72 W)
ź
2
4,8 W/m - UV-C radiation (UV-C STERILON FLOW 144 W)
ź
The irradiation area for direct disinfection from a distance of 1 m is:
2
72 W - 7 m
ź
144 W (one direction) - 7 m
ź
144 W (two direction) - 2 x 7 m
ź
The e ciency of UV-C radiation decreases in inverse proportion to the square of the
as
distance from the light source. Therefore, the closer we place the light source to the
surface requiring disinfection, the more e ective the light will be. The approximate
time for surface disinfection by UV-C STERILON luminaires is 15 minutes. However, this
time may di er from the above as it depends on the power of the lamp used and the
distance of the disinfected surface from the light source. This time also depends on the
ambient temperature and humidity.
Exposure method:
Set the lamp so that the light falls on the surface to be disinfected.
ź
The UV-C light source should be positioned as close as possible to the surface to be
ź
disinfected, and also, so that the scope of covers the largest part of it. Objects
placed in the shade, where the UV-C rays cannot reach, will not be disinfected.
Open the luminaire covers before turning it on.
ź
After turning the luminaire on, you cannot stay in the
ź
room. When turning it on/off, use protective glasses and
wear long-sleeved clothing and gloves.
Approximate exposure time is 15 min.
ź
Note that very long exposure to the UV-C lamp can degrade some elements within
ź
its range, e.g. plastics.
We recommend ventilating the room every 4 operating hours of the device.
ź
We recommend using the product in rooms with
ź
unobstructed ventilation.
Do not use in rooms smaller than 6 m
ź
Disinfection e ectiveness of SARS-CoV-2 (coronavirus causing Covid 19)
There are no published o cial studies confirming the dose of energy that is needed to
neutralize the virus that causes COVID-19 disease. However, its structure is compared
to the "Hepatiti A Virus" virus more resistant to UV-C rays, for which the dose is almost 7
2
mJ/cm . Based on this, the dose needed to neutralise the coronavirus, which causes the
COVID-19 disease is estimated to be approximately 4 mJ / cm . This means that the
lamp emitting 1 W of UV-C radiation at a distance of 1 metre from the surface needs 40
seconds to disnfect 90% of the surface from the virus.
.
Dane techniczne:
PL
2x PL-L 36W, 220-240V, 50-60Hz
Przewód 3m, 3x0,75mm
Przewód 3m, 3x1,5mm
Kl. Ochronności I
Z wtyczką UNI-SCHUCKO / UK
Nie zabezpieczone przed wybuchem
basic
Technische Daten:
DE
2x PL-L 36W, 220-240V, 50-60Hz
3m Leitung, 3x0,75mm
3m Leitung, 3x1,55mm
Schutzklasse I
Mit UNI-SCHUCKO / UK stecker
Nicht explosionsgeschutzt.
UV-C STERILON FLOW 144W - INSTALLATION INSTRUCTION
2
2
2
2
Technical Data:
EN
2x PL-L 36W, 220-240V, 50-60Hz
Cable 3m, 3x0,75mm
Cable 3m, 3x1,5mm
Protection class I
With UNI-SCHUCKO / UK plug
No protection against explosion.
RU
Техническа яхарактеристика
2x PL-L 36W, 220-240V, 50-60Hz
3m Провод , 3x0,75mm
3m Провод , 3x1,55mm
Классзащиты: I
штепсель: UNI-SCHUCKO / UK
Не взрывозащищенный
2
7