6.3 Prime
Prime the pump by pressing Prime
1.
operate at the flow rate and for the duration configured in
the settings menu for the prime feature.
6.4 Flow Calibration
Press Settings
1.
Navigate to the calibration icon
2.
Prepare a device such as a calibration column to measure the total volume of fluid pumped
3.
during calibration. Press the Start / Stop button to start the pump. Allow the pump to run
as long as possible. Calibration accuracy improves with more strokes. Press the Start /
Stop button to stop the pump.
Adjust the total volume using the Up / Down buttons to match the actual volume measured
4.
during the calibration.
1

6.5 Operation Modes

Enhanced control pumps feature multiple operation
1.
modes. Press Mode Select
operation modes.
Manual mode
Analog mode*
Batch mode*
. The pump will
.
. Press Enter
to switch between
Pulse mode*
Cycle timer mode*
Timed event mode*
1
.
2
3
1
7
OPERATION
4
* Enhanced control only