Basic Data

The HLC by DITABIS Thermo Shakers and Block Thermostats are used for tem-
pering and mixing of solutions in closed reaction tubes and plates. Please pay
attention that the target temperature of the samples should not exceed the
boiling point of the solution, which should be heated. Please consider this
when setting the target temperature via the software of the Thermo Shaker
or Block Thermostat. If there are any uncertainties regarding the liquids,
which should be heated or cooled, please contact DITABIS. The HLC by
DITABIS devices are meant for indoor usage only. Please use only HLC by
DITABIS accessories or accessories recommended by HLC by DITABIS.

Safety

The systems comply with the standards and directives mentioned in the
applicable CE declaration.
Please take note of the following safety measures:
Do only connect the system to an earthed mains power
socket of 230 V, 50 Hz.
Do not place hot blocks on inflammable or not heat-resistant
surfaces.
Make sure to only use containers in the block which are
suitable for the desired temperature range.
If liquid gets into the system, immediately pull out the plug
and contact our service department to ensure complete
safety.
Please do not touch or transport hot systems.
Contact with highly flammable fluids must be avoided.
The ventilation slots of the device must not be blocked, to
ensure the proper ventilation at all times.
Damaged mains cable should only be replaced with equal
ones.