Clearaudio WEIGHT WATCHER Bedienungsanleitung Seite 3

Tonabnehmerwaage
Vorschau ausblenden Andere Handbücher für WEIGHT WATCHER:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Clearaudio has developed this digital stylus pressure gauge WEIGHT WATCHER for
accurate stylus pressure adjustment of nearly all phono cartridges. Proper tracking
force is important for keeping the stylus in the groove and creating the optimal
relationsship between the coils attached to the catilever and the magnets in the
cartridge. Today's best phono cartridges demand a specific tracking force
recomended by cartridge manufacturer to optimize trackability and maximize the
sound performance.
Special features of the WEIGHT WATCHER stylus pressure gauge are the automatic
calibration when switched on as well as the and battery protection through automatic
switch off after 30sec.
If the WEIGHT WATCHER is in often use, we recommend to practise a full calibration
as follows:
-Full-calibration - (recommended to be done every three month)
5.
Switch off the WEIGHT WATCHER.
6.
Switch the WEIGHT WATCHER on while pressing the „on / off" Button
once. Immediately after that push the
"Mode" – button until the display shows
"CAL".
7.
Then push the „Mode"-button once again
and the display shows „100.00"
8.
Place the calibrated weight (included in
delivery) in the middle of the surface marked
with „Do not touch or press this surface".
9.
After 5 seconds the display shows „PASS"
and the full calibration is completed.
Please insert the two batteries (type: AAA) as shown in the drawing on the right.
Then you can place the „WEIGHT WATCHER" on the turntable platter surface.
Please proof, that all four small nobs on the bottom of the gauge have contact to the
platter surface. This way it is guaranted that the WEIGHT WATCHER stylus pressure
gauge gives you exact measuring results.
Then press the bottom "on / off" and the WEIGHT WATCHER is starting the
automatic calibration procedure and after 3 sec the Display shows the value "0.0".
Now you can start with the measurement of your stylus presure. Therefore place your
tonarm/cartridge above the „measuring area" marked surface and lower the tonarm
slowly down, so that only the diamond tip touches the measuring area, the cartridge
body should not touch the WEIGHT WATCHER at any point.
Now the WEIGHT WATCHER indicates the absolute stylus pressure stated in
gramms on the display. Follow the manufacturer´s advise to set the correct stylus
pressure for your cartridge.
Downloaded from
www.Manualslib.com
E
- Setup -
manuals search engine
- Attention -
Precise measurements requires that surface is levelled horizontal, when the WEIGHT
WATCHER – gauge is in use.
Never touch, cover or put anything on the surface marked „do not touch or press
this surface" while using the WEIGHT WATCHER to avoid mismeasurements.
Replace the batteries when the WEIGHT WATCHER – gauge cannot be switched
on, or if the measurements results are out of normal range. Can be proofed with the
100gr. calibration weight, included in the delivery.
Inserting the batteries: