Operation
1.
2.
Warning!
Spin balanced loads only! Do not place the tubes asymmetrically nor
load unequal volume tubes. Improper placement of tubes will lead to
insufficient centrifugation and may cause injury or accident.
3.
4.
5.
6.
Downloaded from
www.Manualslib.com
Remove the contents from the package, examining them carefully for
breakage, defects, or missing parts. Save the packaging.
Place Sprout
Plus on a sturdy, level surface at least 12" (30cm)
®
away from the edge of the counter and other moving equipment.
Make sure the power cord is clear of hot surfaces and other hazards.
Prior to every use, inspect unit for cracks, missing pieces, and or
any damage that would result in an impaired performance and notify
Heathrow Scientific BEFORE use.
Verify that the power switch is in the OFF position. Plug the power
adapter cord into the rear of the unit and the power cord into an
approved outlet. Make sure that the power cord connector can be
readily grasped at the rear of the device so that it can be removed to
easily disconnect It from the centrifuge.
Turn power switch ON. If there is a smooth whirring sound
and the unit accelerates with little or no vibration, Sprout
is ready to use. If there are loud or unusual sounds, or excessive
vibration, DO NOT OPERATE. Contact Heathrow Scientific
Technical Support.
On benchtop prepare samples tubes in rack so that tubes are filled to
equal levels. Close tube lids.
To open lid, press down on front latch. The lid will open slightly.
Grasp lid and tilt lid toward rear of unit.