Herunterladen Inhalt Inhalt Diese Seite drucken

Flow FL-100 Bedienungsanleitung Seite 84

Inhaltsverzeichnis

Werbung

18. EMC - Test Results
The device does not emit any radiation intended for
medical purposes, neither hazardous nor potentially
hazardous radiation.
EMISSION LIMITS & MEASUREMENT
PHENOMENON
STANDARD
RADIATED RF
IEC 60601-1-2
EMISSION
CONDUCTED
IEC 60601-1-2
RF EMISSION
VOLTAGE
FLUCTUATION
IEC 60601-1-2
& FLICKER
162
TEST METHOD
TEST LEVELS
30~1000 MHz ; 1~3 GHz
CISPR 11
& 3~6 GHz
ETSI EN 301
(Measurement Uncertainty
489-1
4.24 dB)
CISPR 11
0,15~30 MHz
ETSI EN 301
(Measurement Uncertainty
489-1
3.35 dB)
EUT VALUES & LIMITS
Pst: 0.028 ; Limit 1.00
Plt: 0.028 ; Limit 0.65
IEC 61000-3-3
dc [%]: 0.0000 ; Limit 3.30
dmax [%]: 0.053 ; Limit 4.00
dt [s]: 0.000 ; Limit 0.50
PHENOMENON
ELECTROSTATIC
DISCHARGE
RF
ELECTROMAGNETIC
RESILT
FIELDS
EL FAST TRANSIENTS
COMMON MODE /
PASS
BURSTS
SURGES LINE
PASS
TO LINE
SURGES LINE
TO GROUND
PASS
RF COMMON MODE
VOLTAGE DIPS &
INTERRUPTIONS
IMMUNITY LIMITS & MEASUREMENT
TEST
STANDARD
METHOD
±2 kV, ±4 kV, ± 6 kV, ± 8
IEC 61000-
kV - Contact Discharge
IEC 60601-1-2
4-2
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV - Air Discharge
10 V/m
IEC 61000-
IEC 60601-1-2
80 – 2700 MHz
4-3
80 % AM at 1 kHz
IEC 61000-
± 2 kV for A.C. Power Lines
IEC 60601-1-2
4-4
100 kHz Burst Frequency
IEC
IEC 61000-
± 0.5 kV & ±1 kV At 0°,
60601-1-2
4-5
90°, 180°, 270° ; Repetition
Rate: 60 S
IEC 61000-
± 0.5 kV, ±1 & ±2 kV At 0°,
IEC 60601-1-2
4-5
90°, 180°, 270°
3 & 6 V for A.C. Main Port
; 80%, 1 kHz Amplitude
IEC 61000-
IEC 60601-1-2
Modulation)
4-6
Frequency: 150kHz to
80MHz
0 % UT; 0,5 Cycle At 0°,
45°,90°, 135°,180°, 225°,
IEC 61000-
IEC 60601-1-2
270°, 315°,
4-11
0 % UT; 1.0 Cycle
70 % UT; 25/30 Cycles
TEST LEVELS
RESILT
PASS
PASS
PASS
PASS
PASS
PASS
PASS
163