Herunterladen Inhalt Inhalt Diese Seite drucken

Welch Allyn Health o meter PROPLUS 2500KGWA Bedienungsanleitung Seite 14

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 93
2500KGWA / 2600KGWA CALIBRATION
This scale has been factory calibrated and does not require calibration prior to use.
maintenance, the scale can be calibrated. Calibration of your ProPlus
1.
Make sure the display is turned off, then press and hold the
HOLD/RELEASE button while pressing the ON/OFF or I/O
button.
Using the  keys and/or the numeric keypad, set the calibration
2.
load to at least half the capacity of the scale (for the best results
and greater accuracy, use the full capacity of the scale.
3.
Make sure there is nothing on the scale, then press the
ENTER button.
4.
Please wait 1-3 seconds, without touching or shaking the
scale, to allow the Zero Calibration to finish.
5.
No action required by operator.
6.
Load the scale with the required weight and press the ENTER
button. Please wait 1-3 seconds, without touching or shaking
the scale, to allow the Calibration to finish.
7.
If calibration is successful, the display will show "Factor OK!".
If the calibration process failed, it will show "Factor Fail!".
8.
If successful, the display will begin to reboot.
9.
The display will return to the start up message.
10. Please wait until the scale resumes normal operation and
then remove the weight from the scale.
Operator Action
®
scale is performed using kilos (kg)
13
For
Display
"Enter load weight:
"200.0KG"
"Enter load weight:
"200.0KG"
Zero calibration:
Please clear the scale
Zero calibration:
Please wait
Weight calibration:
Put: XXX.X
Weight calibration
Please wait
Factor Update...
"Factor OK!"
or "Factor Fail!"
Rebooting...
Please wait
Health o meter
ProPlus
Health o meter
ProPlus

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Verwandte Produkte für Welch Allyn Health o meter PROPLUS 2500KGWA

Diese Anleitung auch für:

Health o meter proplus 2600kgwa

Inhaltsverzeichnis