Herunterladen Inhalt Inhalt Diese Seite drucken

human care Silhouette Sling Net Benutzerhandbuch Seite 14

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Slings
Disinfection Procedure
Human Care slings are intended to be used with
one patient at a time. To avoid possible trans-
mission of infectious diseases, it is important to
clean the sling if it has become soiled or if it is to
be used by a different patient.
Approved Disinfectants
Chemical Disinfection: 70% Ethanol, 45% isopropanol or similar
Thermal Disinfection: Wash 65°C (149°F) for a minimum of 10 minutes
Thermal Disinfection: Wash 71°C (160°F) for a minimum of 3 minutes
Note: Add 8 minutes to both thermal disinfection options to ensure heat has penetrated the fabric.
Finish with a rinse cycle.
1
Ref: HSG (95) 18: Hospital Laundry Arrangements for Used and Infected Linen, Department of Health, NHS (National Health Service, England)
1
14
www.humancaregroup.com
1
1
Solutions to Avoid
Steam sterilization
Iodine based disinfectants
Gas sterilization