Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Example:
− writing request of logic window n. 10 with the
logic value '0' (OFF):
MASTER to SLAVE: <0x02> + <0x80> +
<'010'> + <0x31> + <0> + <0x03> +
<'82'>
SLAVE to MASTER: <0x02> + <0x80> +
<0x6> + <0x03> + <'B2'>
When the MESSAGE has a wrong <CRC>, the
SLAVE will respond with the following string:
<STX>+<ADDR>+<NACK>+<ETX>+<CRC>
where <STX>,
<ADDR>,
have the meaning previously discussed, while
<NACK> = 0x15.
When the MESSAGE requires an access to an un-
known window, the SLAVE will respond with the
following string:
<STX>+<ADDR>+<'2'>+<ETX>+<CRC>
where <STX>,
<ADDR>,
have the meaning previously discussed.
When the MESSAGE is a writing command to a
window, but the data is of a wrong type, the SLAVE
will respond with the following string:
<STX>+<ADDR>+<'3'>+<ETX>+<CRC>
where <STX>,
<ADDR>,
have the meaning previously discussed.
When the MESSAGE is a writing command to a
window, but the data has a value out of the admit-
ted range, the SLAVE will respond with the follow-
ing string:
<STX>+<ADDR>+<'4'>+<ETX>+<CRC>
where <STX>,
<ADDR>,
have the meaning previously discussed.
When the MESSAGE is a writing command to a
reading only window, or it is a reading command of
a writing only window, the SLAVE will respond with
the following string:
<STX>+<ADDR>+<'5'>+<ETX>+<CRC>
where <STX>,
<ADDR>,
have the meaning previously discussed.
<ETX> and <CRC>
<ETX> and <CRC>
<ETX> and <CRC>
<ETX> and <CRC>
<ETX> and <CRC>
33
TECHNICAL INFORMATION

Window Meanings

N.
Read/
Data
Write
Type
000
R/W
L
008
R/W
L
100
R/W
L
101
R/W
N
102
R/W
N
103
R/W
N
104
R/W
L
105
R/W
N
108
R/W
N
109
W
L
110
R/W
L
120
R/W
N
Description
Admitted
Values
Start/Stop
Start = 1
(in remote
Stop = 0
mode the
(default = 1)
window is a
read only)
Remote (de-
Remote = 1
fault) or Se-
Serial = 0
rial
(default = 1)
configura-
tion
Soft Start
YES = 1
(write only in
NO = 0
Stop condi-
(default = 1 the
tion)
first time and 0
the following)
Set Point 1
0 = Frequency
type
1 = Current
2 = Time
(default = 0)
Set Point 1
Hz: 150 to 1050
threshold
mA: 0 to 500
9
(expressed
s: 0 to 4x10
in Hz, mA or
(default = 1000)
s)
9
Set Point 1
0 to 4x10
delay: time
(default = 0)
between the
pump start
and the set
point check
(seconds)
Set Point 1
0 = output signal
output sig-
high
nal when
1 = output signal
the meas-
low
ured value
(default = 0)
is greater
than the
threshold
one
Set point 1
0 to 100
hysteresis
(default = 2)
(in % of
threshold)
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
Pump life/
To reset write '1'
cycle time/
cycle num-
ber reset
Interlock
Impulse = 0
type
Continuous = 1
(default = 1)
Rotational
150 to 1050
frequency
(default = 1050)
setting (Hz)
87-900-924-01(B)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis