DEVICE SETUP INSTRUCTIONS
Sterile Adapter System: ACC-GFU-100
SETUP
SET AND PURGE REGULATOR
Set system regulator to operating pressure. If regulator
was set to a higher pressure, purge excess gas by mo-
mentarily connecting regulator hosing to laser gas input,
as shown.
ATTACH FILTER
Attach filter and regulator hosing to laser gas input. Do
not turn on gas flow until fiber is connected to system.
CONNECT FIBER
Connect fiber to laser output, begin gas flow by opening
regulator, and set final regulator pressure if needed.
Never use filter above 100 PSI, which could rupture filter.
The filter is sold sterile for a single use, reuse may pose
WARNING
risk of infection. Do not resterilize this product. Do not
use product if package is damaged.
© OmniGuide, Inc. 2018
SPECIFICATIONS
PARAMETER
VALUE
Maximum Inlet Pressure
100 PSI (7 bar)
Acceptable Gasses
He Medical Grade
Filter Membrane Pore Size
0.2 μm
99.9999% retention when challenged with
Viral Retention
φX174 Bacteriophage per ASTM F2101
Full bacterial retention when challenged with
Bacterial Retention
a minimum of 10 of Brevundimonas Diminu-
ta per cm² of filtration
REORDER INFORMATION
PART NUMBER
DESCRIPTION
ACC-GFU-100
Gas Filter Unit, 10 pack
CALL 1-888-666-4484
FOR SUPPORT, FILTERS, AND PRODUCT INFORMATION
ENGLISH
r1-0168-028-15-00 Rev. 4