To Check The Liquid Oxygen Level - DEHAS EasyMate Reservoir PM2300CE SERIE Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

liquid level
Indicator
Activation
Button

To check the Liquid Oxygen Level:

Push the Activation Button on the liquid level Indicator. The lights will scroll from
bottom to top, then the lights indicating the level of liquid oxygen will remain on for
5 seconds.
100%
75%
50%
25%
0%
100%
75%
50%
25%
0%
100%
75%
50%
25%
0%
Each light represents approximately 12.5% of what the Reservoir
is designed to contain.
EXAMPLE:
This level sensor indicates that Reservoir is
approximately 75% Full of liquid Oxygen.
When ONLY two (2) lights are illuminated, the Reservoir has 25%
of liquid Oxygen; ALMOST EMPTY. Contact your liquid Oxygen
Supplier to fill the Reservoir.
• When ONLY the bottom light is flashing, indicates that the
Reservoir is nearly empty; RESERVOIR NEEDS TO BE FILLED.
• A flashing bottom light for a half second "ON" and a half
second "OFF" for an eight (8) second duration, indicates the
battery needs to be replaced.