Periodical Maintenance Checks - Ca-Mi AS-36BR Handbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 34
GB
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to
remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C
(1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown. Mechanical resistance of the jar is
guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit the
physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
Instruction for disposal Liner Flovac®:
If the device is equipped with disposable collection systems FLOVAC ® carry out the disposal of the bag as follows:
Turn off the Vacuum and remove all the tubes connected to the Liner, giving particular attention to avoid accidental contamination.
Fit the appropriate plugs to the "PATIENT" and "TANDEM" ports, pressing the home firmly, taking care to avoid accidental
contamination. Remove the liner bag from the rigid container and transfer it to the waste disposal area, ensuring that all the
openings are sealed, keeping in mind the product is potentially infectious. This product must be disposed of in accordance with the
current hospital regulations.
The AS-36 BR suction equipment does not need maintenance or lubrication. It is, however to inspect the unit before each use. With
regard to training, given the information contained in the user manual and since it is easy to understand the said device, it doesn't
appear to be necessary. Unpack the instrument and always check integrity of plastic parts and AC/DC switching adapter , feeding
cable, they might have been damaged during previous use. Connect cable to electrical network and turn switch on. Close the
aspirator outlet with your finger and with suction regulator in maximum vacuum position check that the vacuum indicators
reaches – 80kPa (-0.80 bar) minimum (internal battery). Rotate the knob from right to left and check the aspiration regulating
control. The vacuum indicator should go down – 25kPa (-0.25 bar). Verify that loud noises are not present, these can indicate
wrong functioning. The device is protected by a safety fuse (F 10A L 250V) situated in the cigarette lighter cable. When replacing,
always check the type and value as indicated. Internally, the device is protected (see electrical specifications) by two fuses
(T 15A L 125V) that cannot be reached from the outside. Therefore, contact the manufacturer to request the assistance of an
authorized and qualified technician when they need to be replaced. If it's replaced make sure that its replacement is always the
same type and value, as indicated. The device is made up of a lead battery which cannot be accessed by outside. In order to replace
it, consult the technical staff authorised by the manufacturer.
Fault type
1. Red light on
2. No light
3. No aspiration
4. No aspiration
5. The float doesn't close
6. The float doesn't close
7. Low suction
8. No aspiration due to flow
leakage of mucus
9. The Vacuum power on the
patient side is either very
low or absent
10. Noisy
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8
– 9 - 10

PERIODICAL MAINTENANCE CHECKS

Cause
Battery run down
Defective AC/DC adapter or
technical internal problem
Jar Cap badly screwed down
Lid seal not in its seat
If the cap has been washed,
ensure that the float is not
partially detached
The float it's covered by dirty
material
Foam inside the jar
Filter blocked
 Vacuum regulator set to
minimum
 Protection filter blocked or
damaged
 Connection tubes blocked,
kinked or disconnected
 Shut-off valve blocked or
damaged
 Pump motor damaged
Technical internal problem
None of the remedies has
achieved the desired results
Hook up the power cord to the electricity mains,
positioning the equipment power switch on 0.
Contact the technical service
Unscrewed the cap, then rescrew it correctly
Unscrew the cap and insert the seal properly in its seat
Insert the float into it's place
Unscrewed the cap, leave the and put in on autoclave
Fill the jar to 1/3 full of ordinary water
 Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
 Replace the filter
 Replace or reconnect the tubes, check the jar
connections
 Empty the jar, or disconnect the tube from the jar
and unblock the shut-off valve. The unit twill only
work in the upright position
 Contact the technical service
Contcat the technical service
Contact the seller or CA-MI After-sales Assistance
17
Solution
Replace filter
Service