◆ Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15
l/min for the High Flow Blender, and 3 l/min for the Low Flow Blender.
DO NOT obstruct the alarm.
DO NOT use Blender when alarm is sounding.
DO NOT use oil in or around the Blender.
DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender.
DO NOT use near any type of flame or flammable/explosive substances, vapors or atmosphere.
◆ Oxygen Concentration Dial does not rotate 360 degrees. Rotating the dial less than 21% or over
100% oxygen will damage the Blender.
CAUTION
◆ Turn off gas supplies when Air-Oxygen Blender is not in use.
◆ Store the Air-Oxygen Blender in a clean, dry area when not in use.
◆ The Air-Oxygen Blender contains magnetic, ferrous material that may affect the results of an MRI.
◆ Ensure all connections are tight and leak free.
◆ Avoid excessive pressure surges greater than 100 psi (6.9 bar) when pressuring the Blender inlets.
DO NOT steam autoclave.
DO NOT immerse Air-Oxygen Blender into any liquid.
DO NOT gas sterilize with (EtO) Ethylene Oxide.
DO NOT use if dirt or contaminants are present on or around the Blender or connecting devices.
DO NOT smoke in an area where oxygen is being administered.
DO NOT clean with aromatic hydrocarbons.
◆ Inlet pressure of device used in conjunction with Blender must match inlet pressure of Blender.
◆ When using a bottled high pressure gas source, always use a pressure reducing regulator set within
30-75 psi (2.1-5.2 bar).
866.4.MAXTEC
iv
WWW.MAXTEC.COM