EMC data
1. This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be af-
fected by portable and mobile RF communications equipment.
2. Do not use a mobile phone or other devices that emit electromagnetic fields, near the
unit. This may result in incorrect operation of the unit.
3. Caution: This unit has been thoroughly tested and inspected to assure proper perfor-
mance and operation!
4. Caution: this machine should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, this machine should be observed to
verify normal operation in the configuration in which it will be used.
Refer to further guidance below regarding the EMC environment in which the device
should be used.
GUIDANCE AND MANUFACTURE'S DECLARATION – ELECTROMAGNETIC EMISSION
The BW-PFX1 is intended for use in the electromagnetic environment specified below. The customer of
the user of the BW-PFX1 should assure that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emission
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
68
BewellConnect - BW-PFX1 - User Manual - 072017v1
Electromagnetic environment - guidance
The BW-PFX1 use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The BW-PFX1 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
GUIDANCE AND MANUFACTURE'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-PFX1 is intended for use in the electromagnetic environment specified below. The customer or
the user of BW-PFX1 should assure that it is used in such an environment.
EN 60601
Compliance
Immunity test
test level
level
Electrostatic
±6 kV contact
±8 kV contact
discharge (ESD)
±8 kV air
±15 kV air
IEC 61000-4-2
±2 kV for power
Electrical fast
supply lines
±2kV for power
transient/burst
±1 kV for input/
supply lines
IEC 61000-4-4
output lines
± 1 kV line(s) to
Surge
line(s)
±1 kV differen-
IEC 61000-4-5
± 2 kV line(s) to
tial mode
earth
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5 cycle
for 0.5 cycle
Voltage dips,
40% U
40% U
short interrup-
T
(60% dip in U
)
(60% dip in U
tions and volt-
T
for 5 cycles
for 5 cycles
age variations on
70% U
70% U
power supply
T
(30% dip in U
)
(30% dip in U
input lines IEC
T
for 25 cycles
for 25 cycles
61000-4-11
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5 sec
for 5 sec
Power frequency
(50Hz/60Hz)
3 A/m
30 A/m
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
BewellConnect - BW-PFX1 - User Manual - 072017v1
Electromagnetic environment -
guidance
Floors should be wood, concrete or ceramic
tile. If floor are covered with synthetic material,
the relative humidity should be at least 30%.
Main power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
T
)
T
Mains power quality should be that of a typical
T
)
commercial or hospital environment. If the user
T
of the BW-PFX1 requires continued operation
during power mains interruptions, it is recom-
T
)
mended that the BW-PFX1 be powered from an
T
uninterruptible power supply or a battery.
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
69