Dip Switch Set-up

To set-up the KT8 recorder for your specific application,
you might need to change some of the Dip Switches. The
Dip Switches are located on the back of the unit (figure
3). Use a pen or small screw driver to flip the switches.
Remember to install correct chart to match
corresponding Dip Switch setting.
Slide toggle up for on position.
1
2
Dip Switch 7
& 8 are not
active on this
recorder.
Slide toggle down for off position.
Dip Switches for ranges
KT800/01/02/03/05/55/56
DICKSON KT8 Dip Switch Settings:
Up = On; Down = Off
Switches 1 - 6
For Model
7 Day
All Models
24 Hour
All Models
31 Day
KT802/03/55/56
Fahrenheit
All Models
Celsius
All Models
-50° to 50°F/C
All Models
0° to 100°F/C
All Models
0° to 250°F/C
All Models
0° to 500°F/C
KT802/03/55/56
Remember to install the correct chart to match
corresponding switch setting
- Addison, Il
Dickson
KT804
- Addison, Il
Dickson
Rev. 03/08
3
4
5
6
7
8
1 2 3 4 5 6
D
D
U
D
U
U
D
U
U
D
D
D
D
D
U
U
D
U
D D
www.dicksondata.com
www.DicksonData.com
Recording Time
The KT8 recorder has three/two different
recording time options
24 hour
7 day
31 day (KT802/03/05/55)
°F/C
You can record in °F or °C with the KT8 by
using dip switch #3.
°F
°C
Temperature Range
KT800/01/02/03/05/55/56 dip switches
for ranges
0 to +100°F/C
-50 to +50°F/C
0 to +250°F/C
0 to +500°F/C
KT804 dip switches for ranges
-20 to +20°F/C
+20 to +100°F/C
+50 to +150°F/C
0 to +1000°F/C
0 to +2000°F (°F
only)
#1 down
#2 up
#1 down
#2 down
#1 up
#2 up
#3 down
#3 up
#4 down
#5 down
#6 down
#4 down
#5 down
#6 up
#4 down
#5 up
#6 up
#4 up
#5 down
#6 down
#4 down
#5 up
#6 down
#4 down
#5 up
#6 up
#4 up
#5 down
#6 down
#4 down
#5 down
#6 down
#4 down
#5 down
#6 up