Troubleshooting

Error messages

Potential SIP 10 errors are displayed by the photometer.
Table 3 Error messages
Error displayed
Cause
Sipper module not
Connection between
connected.
SIP 10 and photometer
is disrupted.
Please check cable.
Please check sipper
Pump tubing is not
and tube.
correctly inserted.
Resolution
Check the USB cable.
The cable length must not
exceed 1 m (3.3 ft) and the
cable must be directly
connected to the photometer.
Remove other devices that are
connected where necessary.
Loosen the cover and position
again. Where required, run the
pump briefly before the lever is
moved to make sure that the
tube is correctly positioned
around the rollers.

Replacement parts

Designation
SIP 10 Sipper module set for DR 3900 complete with tube
set and 1 inch/1 cm Pour-Thru Cell, EU
SIP 10 Sipper module set for DR 6000 complete with tray,
tube set and 1 inch/1 cm Pour-Thru Cell, EU
SIP 10 Sipper module set DR 6000 complete with tray, tube
set and 1 cm quartz glass Pour-Thru Cell, EU
SIP 10 Sipper module; including pump tubing, EU
USB cable, type AB 1 m (3.3 ft)
Dual Path Length Pour-Thru Cell, plastic, 1 inch/1 cm
UV Pour-Thru Cell, quartz glass, 1 cm
®
Pump tubing (Lagoprene
) pre-assembled with cover and
connections
Complete tube set for DR 3900, including Tygon tubing,
1.70 m (5.6 ft), pump tubing (Lagoprene pre-assembled with
cover and connections), stainless steel tube adapter
Complete tube set for DR 6000 and Dual Path Length Pour-
Thru Cell, including Tygon tubing and Pharmed tubing,
pump tube (Lagoprene pre-installed with cover and
connections), stainless steel tube adapter
Complete tube set for DR 6000 "drinking water application,"
including Pharmed tubing, pump tube (Lagoprene pre-
installed with cover and connections), stainless steel tube
adapter
Tray for storage
Cell compartment cover
User manual, multilingual SIP 10, EU
Information on Tygon tubes, EC no. 1907/2006 (REACH)
Order no.
LQV157.99.10001
LQV157.99.20001
LQV157.99.30001
LQV157.99.00001
LZQ104
LZV876
LZV510
LZV877
LZV875
LZQ102
LZQ100
HTT057
LZQ105
DOC012.98.90328
DOC063.98.90334
English 89

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading