Herunterladen Inhalt Inhalt Diese Seite drucken

Trouble Shooting Guide; Specifications - Mangar International Airflo Plus Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 10
Compressor does not run when the 4 way
hand control buttons are pressed.
Ensure that the Airflo Plus has been
switched on.
Check that there are no kinks in the air hose.
Compressor keeps stopping when trying to
inflate the lifting aid.
Check that there are no kinks in the air hose.
Airflo Plus
Weight:
Output pressure:
Batteries:
Battery Life:
Lifting performance (from full battery):
Environmental protection: IP51 (
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation
Charger: Model number - CA0431
nput:
I
Output:
Double insulated Class II equipment
Simple solutions for everyday independence

Trouble shooting guide

Specifications

5.7Kg (12½ lbs)
0.35 bar (5 psi)
2 x 12v - 2.3Ah (
Maintenance free Valve Regulated Sealed Lead Acid
Approx 1000 charging cycles (
ELK - 3 lifts
Camel - 1 lift (minimum)
Protection against large dust deposits & vertical falling drops of water
AC 100v - 240v, 50-60Hz, 0.3A
DC 30v, 0.45A
Battery State Indicator light flashing 'Red'.
This happens when the compressor's
internal temperature cut-out has tripped.
Leave the Airflo Plus for 60-90 minutes to
cool.
Battery State Indicator light flashing
'Red/Green'.
This happens when the current overload
cut-out has tripped. Leave for 3 seconds to
self reset.
Compressor causes electrical interference
with another piece of electrical equipment.
Move either the compressor or other item
further away from each other until the
interference stops.
2 - 5yrs Depending on usage
Airflo Plus
)
)
)
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis