Herunterladen Inhalt Inhalt Diese Seite drucken

OHAUS MC2000 Bedienungsanleitung Seite 8

Feuchtigkeitsanalysegerät
Vorschau ausblenden Andere Handbücher für MC2000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 37
EN-6
3.3.5
Adjust Mode
Adjust
Up
Down
Category
Enter
4. ADJUST
4.1 Weight Calibration
Before using the analyzer for the first time, we recommend to calibrate it.
1.
Put the instrument on a level surface, remove top cover and empty the
instrument.
2.
Press
button, changes mode to measurement
3.
Long press
4.
Place a 200g calibration mass on the sensor (top of red cone).
Remove the mass when "200" is shown on the display.
5.
6.
Press
button to exit.
4.2 Adjustment
Because of different surroundings and weather, the results will have metrical
discrepancy. The adjustment function will help to reduce the deviation.
1.
Long press
current moisture value. When the moisture value disappears, release
button. The display will show current adjust value, for example "0.0".
2.
Press
and
3.
Press
or
mode.
This function is designed only to adjust the metrical discrepancy
caused by surroundings and weather. Any improper adjustment
should be avoided.
(Long press) Enters adjust mode.
(Short press) Increase adjust value.
(Short press) Decrease adjust value.
(Short press) Save and enter measurement mode.
(Short press) Save and enter measurement mode.
button until "0.0" is flashing.
button to enter the adjust mode. The display will show
buttons to increase or decrease the adjust value.
button to save the value and exit to the measurement
MC2000 Moisture Analyzer Series

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis