Operation
7
Figure 7
2. Push the suction control knob and rotate the suction control knob
until the vacuum gauge indicates the required setting.
g
CAUTION
!
The suction control knob must be completely pushed in to adjust
the vacuum level.
The PTS unit contains a dual spring regulator module to regulate
and adjust suction which is precise in the critical care range (0-200
mmHg/26.7 kPa) and quickly moves to full wall vacuum for resus-
citation. Unlike some other Ohio Medical vacuum regulators, the
PTS requires only two turns from zero to full wall vacuum.
PTS units are shipped from Ohio Medical set at 100 mmHg and
require 1-1/2 turns to maximum.
Failure to follow these cautions may damage the vacuum regulator.
Pre-Use Checkout Procedure
!
WARNING
The Pre-Use Checkout Procedure must be performed
before using the equipment on each patient. If the
regulator fails any part of the Pre-Use Checkout Pro-
cedure, it must be removed from service and repaired
by qualified service personnel.
All tests must be performed with supply vacuum of 500 mmHg (67 kPa)
minimum.
1. Turn the mode selector switch to O (OFF).
Push and rotate the suction control knob one full turn clockwise
(increase).
Release. The gauge needle should not move.
2. Move the mode selector switch to I (ON). The gauge should indicate
vacuum.
6
8700-0007-000
6