Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

RIGOL

AC Current Gain

Test
[1]
Range
Signal
1 mA
20 mA
20.0000 mA
20 mA
20 mA
20 mA
10 mA
200 mA
200 mA
200.000 mA
200 mA
200 mA
100 mA
2 A
2.00000 A
2 A
2 A
2 A
500 mA
10 A
[2]
10 A
10.0000 A
10 A
10 A
[1]
Note
: The reading rate is set to "Slow".
[2]
Note
: For continuous currents greater than DC 7A or AC RMS 7A, close the circuit for 30 seconds
and then open the circuit for 30 seconds.
[3]
Note
: Q marks the optional quick test points.
4
Input
Quick
Frequency
Test
[3]
1 kHz
Q
20 Hz
45 Hz
2 kHz
Q
10 kHz
1 kHz
20 Hz
Q
45 Hz
Q
2 kHz
Q
10 kHz
Q
1 kHz
20 Hz
45 Hz
2 kHz
Q
10 kHz
1 kHz
20 Hz
45 Hz
2 kHz
Q
5 kHz
DM3058/DM3058E Performance Verification Guid
Appendix Test Record Form
1-year Test Limit
Upper Limit
Lower Limit
1.0450 mA
0.9550 mA
20.3200 mA
19.6800 mA
20.1200 mA
19.8800 mA
20.1200 mA
19.8800 mA
20.5400 mA
19.4600 mA
10.430 mA
9.570 mA
203.200 mA
196.800 mA
200.800 mA
199.200 mA
200.800 mA
199.200 mA
205.400 mA
194.600 mA
0.10650 A
0.09350 A
2.03400 A
1.96600 A
2.01400 A
1.98600 A
2.01400 A
1.98600 A
2.05400 A
1.94600 A
0.5275 A
0.4725 A
10.1650 A
9.8350 A
10.0650 A
9.9350 A
10.0650 A
9.9350 A
10.2700 A
9.7300 A
Test
Result

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Dm3058e

Inhaltsverzeichnis