Chapter 2 Performance Verification Test
10.0000 A
10.0000 A
[3]
Resistance
200.000 Ω
2.00000 kΩ
20.0000 kΩ
200.000 kΩ
2.00000 MΩ
10.0000 MΩ
100.000 MΩ
[1]
Note
: The reading rate is set to "Slow".
[2]
Note
: For continuous currents greater than DC 7A or AC RMS 7A, close the circuit for 30 s and
then open the circuit for 30 s.
[3]
Note
: Specifications are for the 4-wire resistance function or the 2-wire resistance function using
"REL" operation. If the 2-wire resistance function is used without "REL" operation, add 0.2 Ω
additional errors.
[4]
Note
: Use the 2-wire resistance function.
[5]
Note
: Q marks the optional quick test points.
2. The AC voltage gain verification procedures
(1) Make sure you have carefully read the "Test Conditions".
(2) Connect the input terminals of the multimeter with the calibrator.
(3) Test the ranges specified in Table 2-3 one by one. Set the reading rate to
"Slow" and turn off all the math operations.
(4) Input the signals specified in Table 2-3 using the calibrator and compare the
test results with the test limits in the table (make sure the output of the
calibrator is built adequately).
Table 2-3 AC Voltage Gain Verification Table
[1]
Range
Signal
10 mV
200.000 mV
200 mV
200 mV
200 mV
200.000 mV
200 mV
200 mV
100 mV
2.00000 V
20.0000 V
200.000 V
DM3058/DM3058E Performance Verification Guide
10 A
-10 A
200 Ω
2 kΩ
20 kΩ
200 kΩ
2 MΩ
10 MΩ
[4]
100 MΩ
Test
Input
Frequency
100 kHz
20 Hz
45 Hz
20 kHz
50 kHz
100 kHz
100 kHz
2 V
20 Hz
2 V
45 Hz
2 V
20 kHz
2 V
50 kHz
2 V
100 kHz
1 V
100 kHz
20 V
20 Hz
20 V
45 Hz
20 V
20 kHz
20 V
50 kHz
20 V
100 kHz
10 V
100 kHz
10.0260 A
-9.9740 A
200.070 Ω
Q
2.00046 kΩ
20.0046 kΩ
200.046 kΩ
Q
2.00088 MΩ
10.0253 MΩ
101.754 MΩ
1-year Test Limit
Quick
Test
Upper Limit
[2]
Q
10.660 mV
203.200 mV
200.500 mV
Q
200.500 mV
202.100 mV
206.100 mV
0.10660 V
Q
2.03200 V
Q
2.00500 V
Q
2.00500 V
Q
2.02100 V
Q
2.06100 V
1.0660 V
20.3200 V
20.0500 V
Q
20.0500 V
20.2100 V
20.6100 V
10.660 V
RIGOL
9.9740 A
-10.0260 A
199.930 Ω
1.99954 kΩ
19.9954 kΩ
199.954 kΩ
1.99912 MΩ
9.9747 MΩ
98.246 MΩ
Lower Limit
9.340 mV
196.800 mV
199.500 mV
199.500 mV
197.900 mV
193.900 mV
0.09340 V
1.96800 V
1.99500 V
1.99500 V
1.97900 V
1.93900 V
0.9340 V
19.6800 V
19.9500 V
19.9500 V
19.7900 V
19.3900 V
9.340 V