Operation

Location

Select a location that allows for 3" clearance behind the
power supply, and comfortable reach of the front panel
controls and cell connections. Do not block the vented
area of the case - on the front bottom of the unit, or the
fan area at the rear. Connect the power supply to a 3-
prong grounded AC outlet, using the AC cord provided
with the unit only. Connect the electrophoresis apparatus
to the power supply, making sure to match the red posi-
tive lead to the red positive jack, and the black negative
lead to the black negative jack. Similarly, connect a sec-
ond or more electrophoresis apparatus to the power sup-
ply if you plan to run more than one in parallel. Attach a
temperature probe to the FB3000 rear located connector
and the electrophoresis apparatus if you intend to perform
temperature-controlled runs. Note that only one elec-
trophoresis apparatus can be controlled when using tem-
perature control. Power the unit on using the AC power
switch located at the rear of the unit, next to the AC line
cord entry.
Figure 1. Front Panel Controls
Black negative jacks
Function keys
Black negative jacks
Exit key
Clear key
Red positive jacks
Red positive jacks
Run key
Pause key
Stop key
gray ON/OFF key