Herunterladen Inhalt Inhalt Diese Seite drucken

Anhang; Immersion Testing Protocol - Dego SpiroLab II Handbuch

Praxis-/klinik-spirometer
Inhaltsverzeichnis

Werbung

Anhang

Immersion Testing Protocol

Date of submission to DuPont:
Supplier/Manufacturer:
Research
Testing Protocol Number:
EQUIPMENT / PARTS SUBMITTED
PART NO.
DESCRIPTION
(DUPONT
)
1, 2 & 3
Turbine Flow Meter
TEST REGIME
IMMERSION MEDIUM
IMMERSION TIME
RINSE MEDIUM
RINSE TIME
FREQUENCY OF MEDIA REPLACEMENT
NUMBER OF CYCLES
FREQUENCY OF AND RESPONSIBILITY
FOR INSPECTION
TEST PROTOCOL APPROVAL.
th
4
May 2006
MIR - Medical International
040506
Geräte-Handbuch zum Spirometer SpiroLab II
Process Owner
Issue Number
Issue date
Review due
Approved by
MATERIAL OF CONSTRUCTION (IF
KNOWN)
Polycarbonate (Makrolon 2805) +
magnetic tape
Perasafe Powder @ 16.2 g/l in tap water
10 minutes
Tap water
15 seconds
24 hours or after 100 immersions
Initial 10 cycles. If no apparent incompatibility
a further 90 cycles to be completed (100
total).
DuPont to inspect after 10 cycles. Test
pieces to be rinsed with distilled water,
allowed to dry and returned to manufacturer
for final evaluation after 100 immersion
cycles.
Chief Chemist
3
4.3.05
4.3.07
Tony Shaddock