Quick Start Guide for Model AX300-I Portable Oxygen Analyzer
and MX300-I Portable Oxygen Monitor
The Quick Start Guide describes the steps to successfully start and
calibrate the instrument. This guide covers the AX/MX 300-I series
of analyzers and monitors. Before placing the instrument in service
please read the users manual that accompanies the device.
Components
The AX/MX 300-I Portable Oxygen Monitor includes:
Sensor & Diverter
Sensor Cable
T-Adapter
V mount Adapter
AA Batteries
Instruction Manual
Set Up
Set up is simple:
1. Connect the coiled sensor cable to the
monitor and tighten the capture nut.
2. Connect the other end to the sensor
and tighten the capture nut.
3. Install 3 AA alkaline batteries.
Calibration
Note: The AX300-I and MX300-I can only be
calibrated using 100% oxygen or room air (20.9%).
Improper calibration or use of other gas
concentrations will activate the
Sensor
Connection
SENSOR indicator.
Monitor
Connection

Battery Installation

1. Hold the instrument face down.
2. Using a coin. Gently twist to open
the hinged battery compartment door
latch.
3. Lift up door near latch and pull out.
4. Install 3 AA alkaline batteries
observing proper polarity.
Note: Depressing keys while installing
batteries may activate an error code.To
reset remove and replace battery.
Note: Use only alkaline batteries
Micro-Fuel Cell
Sensing
Surface
Flow
Diverter
Tee
Adapter
1