IMPORTANT
Coolant and dirt are prevented from entering
the OMM by the cable sealing gland.
The OMM cable is protected against physical
damage by fitting either flexible or rigid conduit.
The OMM conduit adaptor supplied by
Renishaw accepts both types of conduit,
provided the correct components are selected
during assembly.
Before fitting nut C, grease the adaptor thread
with a general purpose grease
e.g. Shell Alvania grease.
When tightening or loosening nut C onto
conduit, ensure that torque is only applied
between B and C.
If during the fitting of component A, nut B
or nut C become loose, the gland will be
ineffective.
WARNING
Failure to adequately protect the cable can
result in system failure due to either cable
damage or coolant ingress through cores
into the OMM.
Failure due to inadequate cable protection
will invalidate the warranty.

OMM CABLE SEALING

{
Conduit
Component - A
adaptor
CONDUIT ADAPTOR
If Nut - C is removed, fit a new olive before
replacing Nut - C. The olive is available as
part of Thomas & Betts Gland (Part No. 9330).
PG TYPE CONDUIT
Where PG type conduit is to be used, an
optional adaptor is available from Renishaw.
The adaptor will convert the cable gland to
a PG thread. (Adaptor Part No. M-2008-0189).