Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration With Calibration Solution - Knick Stratos PROFIBUS 2221X Cond Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Calibration

Calibration with Calibration Solution

Display
Action
Press cal,
enter mode code 1100.
Select using ,
enter number using ,
press enter to proceed.
Welcome (3 sec)
Immerse the sensor in the
calibration solution.
The upper display shows the
cell constant, the lower display
shows the measured conductiv-
ity value.
Read off the conductivity value
for the temperature displayed
from the table of the calibration
solution used.
Change the cell constant,
until the conductivity value
(lower display) corresponds to
the value specifi ed in the table.
Press enter to confi rm the cell
constant.
Press enter to exit the HOLD
mode.
Remark
If an invalid code is en-
tered, the device returns
to measuring mode.
Device in HOLD mode,
Sensoface inactive.
When there has not been
an entry for 6 sec, the
lower display alternately
shows the conductivity
and temperature value.
113

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis