4
Mounting the Mixer
You mount the mixer with 2 Allen screws. The screws are included with
the accessory kit.
Tools
Allen screwdriver, 2 mm
torque wrench, 2.5 mm
Hinweis
Process and Figure
1. Attach the mixer and fix it with the screws
2. Using the screwdriver, fasten the screws.
3. Fasten the screw fittings
Afterwards take the device into operation.
Mounting the Mixer
The biocompatible mixer can be used with stainless
steel or PEEK fittings for the capillaries. In case you
use stainless steel, the sealing ring must be made
from PEEK.
CAUTION!
Damage to components caused by overtightened fit-
tings!
Note the torque of the fittings:
- For stainless-steel fittings: 5 Nm
- For PEEK fittings: maximum 1 Nm or finger-tight
and an additional quarter of a turn with appropriate
pliers.
2
Mixer Supplement V6730, version 1.0
together with the capillaries.
1
2
.
1