Unplug the device before cleaning.
Cleaning with water
Wash all nebuliser components (except air hose) under warm tap water (max 60 °C) for about 5 minutes adding if necessary
a small quantity of detergent following dosage and use limitations as provided by detergent manufacturer. Rinse thoroughly
making sure that all deposits are washed away and leave to dry.
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use limitations
as provided by disinfec tant manufacturer. Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and ma-sks) can be heat steam sterilized up to 121 °C (20 min.) or 134 °C (7 min.).
EN554/ISO11134.The sterilization packaging must conform to EN868/ISO11607 and be suitable for steam sterilization.
After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when compo-nents are still warm.

MAINTENANCE, CARE, AND SERVICE

Order all spare parts from your dealer or pharmacis.
It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization cycles. Check
the filter continually for cleanliness and repla-ce it if dirty, or after a maximum of 3 months use. Spare filters are provided with
the device. To replace the filter, open the air filter compartment 4 marked «Filter» and change the filter 11

MALFUNCTIONS AND ACTIONS TO TAKE

The device cannot be switched on
Ensure the power cord 2 is correctly plugged into the socket. Ensure the ON/OFF switch 3 is in the position «I». The device
has automatically switched off due to overheating. Wait until it has cooled down and try again.
The nebuliser functions poorly or not at all
Ensure the air hose 6 is correctly connected at both ends. Ensure the air hose is not squashed, bent, dirty or blocked. If necessa-
ry, replace with a new one. Ensure the nebuliser 5 is correctly assembled 12. Ensure the required medication has been added.

TECHNICAL DATA

Piston electrocompressor with thermal cut
Nominal voltage: see data label
Dimensions: 130x190x100 mm
Weight: 1.1 Kg approx.
Temporary use: 20/40 min I/O
MDD 93/42/EEC - Class of risk IIa
Maximum pressure: 185 kPa
Air flow: 10 Lt/min
Noise level: 58 dBA (EN13544-1)
Neb rate: 0.35 ml/min (con soluzione NaCl0.9%)
Operating pressure: 60 kPa (0.6bar)
Operating flow: 5 l/min
MMAD: 2.61 μm GSD: 2.75 Aerosol Output: 143 μl Output Rate: 87 μl/min