Maintenance

®
Sprout
is normally maintenance-free. Clean unit only when it is not
plugged into an electrical outlet. When necessary, the
housing and rotor can be wiped using a damp cloth and a mild,
non-corrosive detergent.
Warning!
Do not use any solvent on the unit that may attack plastic or cause cracks
in the rotor. Ensure the rotor is protected from corrosion and mechanical
damage. The rotor must be cleaned with a neutral cleaning liquid.
Notes
—Ensure all parts are throughly dry prior to operation.
Trouble Shooting
Trouble
Not working
— Cord Loose
No light
— Front ON/OFF switch
— Lid not closed
Excessive
— Rotor not installed correctly
vibration
— Rotor not balanced correctly
Noise -
— Lid not fully closed
clicking
sound
Rotor falls
— Rotor not attached
off
Possible Cause
Solution
— Check Cord
— Press front
ON/OFF switch
— Close Lid
— Reinstall rotor
— Check for tubes
of equal weight
and location
(see Operation)
— Close Lid
engaging silicon
tab with cover
— Reattach rotor
(See Rotor
Replacement)
9