Appendix A Installation of the equipment

A.1
Attaching the headrest and the power supply
to the instrument table HSM-801 or HSM-901
The operating instructions of the concerning instrument table
should be observed!
A.2
Attaching the headrest and the power supply
to tables and units of other manufacturers
1. Fix the power supply with 4 screws (73).
2. Remove the knurled screws (70) and the cable plate (68).
3. Fix the ground lead (67) with the screw (69) to the unit.
Tables of other manufacturers: connect the ground lead only
if necessary!
4. Connect the four pole electrical plug connector (71) between
the headrest and the power supply.
5. Install the headrest in the required position on the adapter
piece of the table top by means of both Allan screws (76).
6. The cable plate (68) should be installed in such a manner
that the plug connection and cable are affixed.
7. Mount the handles (77) and connect the main lamp lead (75)
to the lamp housing.
8. Screw the angled plug (74) up tight to the plug socket (78).
9. Attach the potentiometer cable with the cable junction box
(79) on the right side of the rail cover by using the center
groove.
10. Connect the power cable (72).
CAUTION
• Only qualified and trained personnel may
work on cables and components being con-
nected to the power supply.
• To prevent short circuits, do not crimp or
squeeze the electric wires!
72
73
74
75
76
77
78
79
richtig / juste / right
falsch / faux / wrong
1500.7200631.04010
67
68
69
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
70
71