5.0 Cleaning/Disinfection
5.1 General information on cleaning and disinfection
Before cleaning
Medical devices must always offer a maximum in safety and function.
The described measures to clean and to
disinfect do not rule out regulations valid
for the particular company!
●
In chapter 5.3 Recommended disinfec-
tants" all surface and instrument disin-
tectants are stated which are suitable
for disinfection.
5.2 Cleaning the unit's surfaces and upholstery
●
The surfaces of the ATMOS Chair 21 D are resistant against all the recommended surface
disinfectants stated in chapter 5.3 and 5.4. Nevertheless after any length of time discoloura-
tions could possibly develop.
●
Substances such as blood need to be removed immediately to prevent stains on the cover
material of the upholstery.
●
The unit itself can be wiped off with a moist (not wet) cloth.
Do not use any abrasive cleaning agents. For the upholstery usual dry foam may be
!
used.
Treatment with a commonly used care product for artifi cial leather is recommended
once a week in order to keep the upholstery soft and smooth.
Always observe the concentration speci-
fi cations and instructions by the respective
manufacturer.
●
Do NOT use
- disinfectants containing concentrated
organic or anorganic acids or bases,
since these may cause corrosion
damages.
- disinfectants containing chloramides,
phenol derivatives or anionic tensides,
since these may cause stress cracks in
the plastics used.