CAUSE
9.2 Warning: OUT OF SPEC
A
humidity
range
>
was given, or < 10% RH was selected.
9.3 Warning: WATER LEVEL HIGH
The maximum filling level with distilled
(deionized) water was exceeded.
9.4 Warning: WATER LEVEL LOW
The minimum filling level was not
reached.
9.5 Fault message: heat defect
The HUMOR does not achieve its
operating temperature.
9.6 Fault message: pressure excess
A humidity - desired value < 8% RH
was selected. The humidity - generator
can be damaged by the high pressure.
The indicated accuracy of the measu-
rements can no longer be guaranteed.
9.7 Humidity - display is not correct
Incorrect display, e.g. due to an error in
the electronics.
29
95%
RH
HUMOR is operated out of specified
range.
The water must be emptied again until
the fault message disappears.
Distilled (deionized) water must be
filled.
Contact the customer service depart-
ment of E+E Elektronik ® Ges.m.b.H.
Contact the customer service depart-
ment of E+E Elektronik ® Ges.m.b.H.
Test the HUMOR.
Turn the controller all the way to the left
and
close
After a stabilization time of 2h, the
display should show 100 ± 2% RH.
If this is not the case, contact the
customer
E+E Elektronik ® Ges.m.b.H.
SOLUTION
the
compressed
air.
service
department
english
of