For safety reasons the quartz
radiator is equipped with a final
switch and only works in swivelled
in position after having started a
programme.
If the quartz radiator is swivelled
outwards after having started a
programme, the heating process
will be interrupted automatically.
The indication in the display (a)
will stop and show the remain-
ing time. By swivelling the quartz
radiator back in the programme
will be re-activated.
Attention! The swivelling hous-
ing of the heating element
reaches high temperatures.
8.4 Thermoforming
After the heating time has elapsed there
will be an acoustic signal; the heat-
ing time in the display will be 00:00.
Heating and LED illumination will be
switched off automatically.Pull the heat-
ing (4) with the right hand at the knob
rod (5) as far as it will go to the front
and hold it tight.
Then push the push button (8) with the
left hand at the left side of the housing.
The piston will move downwards and
close the pressure chamber; an acoustic
signal will be heard.
Info
The closing pressure of the down-
wards moving piston is always
higher than the forming pressure!
38
Therefore it must not be touched –
even not for short moments. After
the adjusted time has elapsed the
indicator (a) in the display will
stop and count the duration after
termination of the heating time.
This process may be stopped by
abort (long pressing of button 8)
or by successive thermoforming.
The downwards moving piston is
equipped with a dual electronic
safety feature and can only be acti-
vated, if the unit is operated with
two hands at the same time (two-
hand control). Always make sure
that no objects are in the contact
area between piston and F insert
while the piston is closing!
Hold both buttons until the pressure bar
in the display (g) shows the adjusted
minimum of 29 psi (2 bars). Then
release the push button (8) and finally
the knob rod of the heating (5). After
having reached the 29 psi (2 bar) mark
the pressure will automatically rise up
to the pre-adjustment.
Info
will show the last stored values.
For safety reasons the process of
In order to start the thermofor-
an unwanted downwards move-
ming process newly the process
ment of the piston or an interlock
can be repeated immediately, as
of the escapement can be stopped
described in this chapter.
by the electronics:
a) The process of moving the pis-
The adjusted pressure will be verified by
ton downwards is carried out too
the electronics and regulated, if neces-
slowly (max. 3 seconds between
sary. The display shows the adjusted
action right hand and action left
cooling time (c) which then counts
hand) or
down to 00:00.
b) The two-hand control is aborted
before reaching the minimum
pressure.
In both cases there will be 5 short
acoustic warnings, the piston will
be ventilated and remove to its
original position. The display

8.5 Opening of the pressure chamber

In order to attain a complete cool down
Directly before releasing the pressure
of the thermoforming blank the valve
inside the piston the ventilation valve
at the tension ring (12) is to be opened
at the F insert (6) should be opened.
slightly after 2/3 of the adjusted cooling
Push the push button (8) for approx.
time. Through this the warm air inside
two seconds; the pressure inside the
the piston will be released and replaced
piston will be released. After complete
by cool air due to the pressure regula-
ventilation the piston will remove to its
tion of the electronics. After the cool-
original position. Swivel out the plate
ing time has elapsed there will be an
reception (11) at the corresponding
acoustic signal (indication in the display
handle (10) and remove tension ring
(c) will blink).
and thermoformed result.
The display (2) will show the last used
programme; the scanning device (1) is
ready for a new thermoforming pro-
cess.