Printer settings:
DIP switch is used for setting the printer's interface, DIP switch is located on the
right side inside the battery compartment.
IMPORTANT:
1.
In the setting number 2 the setting of IrDA has to be ON and Bluetooth OFF,
otherwise the IrDA connection does not function (Bluetooth is optional and not
available with this printer model).
2.
Do not pull the printing paper forcibly during the printing process. This behavior
damages the printer's head and makes the printing unstable.
There is no "automatic power off". So after finishing the printer operation you
3.
have to turn off the power of the printer. If this is not done, the batteries will be
discharged soon!
Conducting a printer test:
1.
Check if there are 4 batteries inside the battery compartment and if the battery
poles are correct (see chapter „Battery change").
2.
Check if the paper role is inserted correctly (see chapter „Paper change").
3.
Press the paper feed button on front of the printer and additionally the start
button.
4.
The printer starts a test print.
Conducting a print:
1.
Check the batteries and the paper role.
2.
Press the start button on the left side of the printer.
3.
The printer is ready as soon as the green LED light on front of the printer is
shining.
4.
Take the ACE ALP-1 med Breathalyzer, press the start button and choose the the
„General settings" in the menu. Choose „Printer" and then the kind of connection
you want to use. With this printer model you can choose an IrDA connection or a
connection via cable (this cable is part of the scope of delivery, Bluetooth is
optional and not available for this printer-model).
5.
By using the IrDA connection you can only print out the data of the last measure-
ment done with the ACE ALP-1 med breathalyzer. If you choose to print by using a
cable connection between the breathalyzer and the printer you have got 3 variants
of selectable data:
a.
„Last data": If you select this variant on the display of the ACE ALP-1 med breat-
halyzer, only the last measured test result is printed (shortly press the middle button
DIP S/W No
Define
1
OFF
Normal Print
ON
Reverse color print available
OFF
2
IrDA transmitting
(Default for ALP-1)
ON
Bluetooth transmitting (optional)
OFF
3
Normal
ON
Printed data delete available
(optional)
4
OFF
Normal
ON
Printed data storage available
(optional)
on the front of the breathalyzer and the print is started).
b.
„Selected data": If you select this variant on the breathalyzer, you can use the
arrow keys on the right and left of the device front of the alcohol tester to select
a very specific measuring result (1 measurement result of all results stored in the
alcohol tester). Use the left and right arrow buttons on the front of the alcohol tester.
Pressing the middle button confirms the selected data record and it is printed
immediately.
c.
„Range (? -?)": If you select this variant on the breathalyzer, you can select several
successive test results from a range you specify. Press the left and then the middle
button on the breathalyzer. The left number on the alcohol tester display can now
be changed with the two arrow keys up / down. Press the middle button to confirm.
Now you can edit the right number on the alcohol tester display (using the arrow
keys again). To confirm, press the middle button again, then press the left button.
The display jumps to „Next" and when you now press the middle button, the print
is started.
6.
To start the printing via an IrDA connection IrDA please hold the IrDA window on
front of the breathalyzer`s cover near to the IrDA window of the printer and press
the start button of the breathalyzer. As soon as this is done the printer will start
printing the last measurement done with the ACE ALP-1 med.
Technical data:
Printing
Direct Line Thermal
Print Speed (LPS)
50 mm/sec. (Maximum)
Resolution (DPI)
200 (8 dot/mm)
Font size
12x24 / 9x24 (Eng.)
Column
24 / 32 / 36 / 42 Columns
Language
English
Barcode
UPC-A, UPC-E, EAN-13, EAN-8, CODE 39, CODE93,
CODE128, ITF, CODEBAR
Emulation
EPSON / CITIZEN
Connection possibilities
Cable RS-232C / IrDA / USB (optional) / Bluetooth
(optional)
Data buffering
16 KB
Paper
64 mm x 133 mm x 33 mm (W x L x H)
Power supply
4 pieces AA battery (1.5 V / Alkaline) or 7,5 VDC 2A
Working temperature
-10 °C up to 50 °C
Weight
310g (including batteries and paper role)
Dimensions
76 mm x 131 mm x 51 mm (W x L x H)