Herunterladen Inhalt Inhalt Diese Seite drucken

Operation Instruction; Cleaning; Maintenance - Omnibus AIR NEBULIZER Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation instruction

1.Place compressor on a stable, strudy and flat surface, so that the unit can be easily reached when you are seated.
2.Pull the power cord from the base of the unit, make sure that the device is switched off (position O of the power switch ).
3.Plug power cord into the wall-socket.
4.Connect one end of the air tube into the air outlet connector on a nebulizer's body.
5.Unscrew the nebulizer cup (medicine container), check if the SPRAYING NOZZLE is inside. Pour the measure dose of
medicine, prescribed by the doctor.
The sign 6 cc on the container points out the maximum container`s capacity. Do not pour more than maximum level of
medication solution. Tighten the container.
WITHOUT THE THE SPRAYING NOZZLE the NEBULIZER WILL NOT SPRAY A MEDICINE!
6.Put the mouthpiece or the mask on the medicine's container (nebulizer cup).
7.Put the second end of the air tube on the bottom part of the nebulizer cup, from the bottom. Sit down.
8.Switch on the device (position I of the power switch).
9.Put the mouthpiece between teeth, clench lips around the mouthpiece and breath deeply with a mouth.
10.When you use the mask, close attention to suit it to your face. Breath slowly and deepley, doing short breaks after
each inspiration and doing fast expiration.
Inhalations should be conducted on the pitting position!
After finishing operations, the device should be switched off and taken out the plug from the socket.

Cleaning

1. Turn the power off and unplug from the wall outlet .
2. Remove the tubing from the air inlet.
3. TO CLEAN: Disassemble mouthpiece, nosepiece, nebulizer cup, chamber and nozzle, and wash this items in hot
water with dish-washing detergent.Rinse these items thoroughly to remove the detergent and let air dry.
4. TO DISINFECT: Mix one part of white vinegar with 3 parts hot water in a clean container. Submerge mouthpiece,
nebulizer cap, chamber and baffle for half an hour in the solution. Remove from solution and let air dry.
5. There is no need to clean the tubing. If necessary wipe the surface regularly.
WARNING: The nebulizer cup (medicine container) should be changed if it got dirty or plugged.
WARNING: Do not boil, do not scald out the medicine container and the other accessories.
It is recommended to replace accessories every 6 months or 3 months- if used intensively.

Maintenance

1.
The air filter in the compressor should be changed if it only starts to darken. In order to do that raise up the filter's
cover, take out the used filter, put inside the new filter and shut the cover.Use only original filters and accessories!
Additional filters are available at the Seller or on the website:
2.
Repairs can be made only by the authoised service. The service in Poland: Omnibus Medical Equipment, 9
Grunwaldzka St., 82-500 Kwidzyn,
WARNING: The nebulizer can be damaged, if it is used with the dirty or unoriginal filter.
DO NOT USE WITHOUT THE AIR FILTER.
DO NOT LOSE THE COVER PLATE OF THE NOZZLE.
www.omnibus-med.com.pl
www.sklep-omnibus.pl
GB

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Omnibus AIR NEBULIZER

Diese Anleitung auch für:

Br-cn133

Inhaltsverzeichnis