E02: Oxygen measurement out of range Low (≤-2.0% calculated by algorithm).
Recommended Action: Verify that the UltraMaxO
(Concentrator or Calibration Verification mode). If error code repeats; perform a
calibration verification per section 2.3 of this manual. If error code repeats again;
contact Maxtec Customer Service.
E03: Device memory corrupt or missing. Recommended Action: Return the
UltraMaxO
to the manufacturer for factory repair.
2
E04: Signal reading not stable. Recommended Action: Return the UltraMaxO
manufacturer for factory repair.
E05: Pressure measurement out of Range Hi (≥50 PSI). Recommended Action: Check the
pressure on a known gas source pressure. If error code repeats; contact Maxtec Customer Service.
E06: Outside of operating temperature Hi (≥40° C). Recommended action:
The UltraMaxO
is too hot, cool the device closer to room temperature before use.
2
E07: Outside of operating temperature Low (≤15° C). Recommended action:
The UltraMaxO
is too cold, warm the device closer to room temperature before use.
2
E08: Device self check found error. Recommended Action: Remove and replace the
batteries. If error code repeats; return the UltraMaxO

5.0 CHANGING THE BATTERIES

Batteries should be changed by service personnel. Use only brand name batteries.
Replace with two AA batteries and insert per orientation marked on the device.
Batteries should be changed when the
until the batteries are changed. If the battery power level is too low the UltraMaxO
not power on until the batteries are changed.

5.1 Battery Replacement Procedure

1. Using a #1 Phillips screwdriver loosen the battery door screw and remove the battery door.
2. Remove the batteries.
3. Insert new batteries ensuring correct polarity. DO NOT use rechargeable batteries.
4. Replace the battery door and tighten the battery door screw.
5. If the UltraMaxO
correctly and that the batteries are fresh.

6.0 CLEANING AND MAINTENANCE

Use caution to prevent any fluid from entering the UltraMaxO
DO NOT soak or immerse the UltraMaxO
DO NOT autoclave or expose the UltraMaxO

6.1 Cleaning

Wipe down the exterior surfaces of the UltraMaxO
dish soap (pH 6-8).
does not power on when done verify the batteries are installed
2
866.4.Maxtec www.maxtec.com
866.4.Maxtec
is being used in the correct mode
2
to the manufacturer for factory repair.
2
icon illuminates. The icon will remain lit
BAT
in fluid.
2
to ethylene oxide sterilization.
2
with a moist cloth and mild hand or
2
www.maxtecinc.com
to the
2
.
2
will