A2 Electromagnetic Immunity - For Home Healthcare Environment
EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The VP-M10/IN-200 Vapo is intended for use in the electromagnetic
environment specified below; The customer or the user of the VP-M10/IN-200
Vapo should assure that it is used in such and environment.
Immunity
IEC 60601
test
test level
Electrostatic
±8kV contact
discharge
±2kV, ±4 kV,
(ESD) IEC
±8kV, ±15kV
61000-4-2
air
Electrical fast
±2KV for
transient/
power supply
burst IEC
lines
61000-4-4
±1kV Leitung
Surge IEC
zu Leitung
61000-4-5
±2kV Leitung
zu Erde
Power
frequency
(50/60Hz)
30A/m
magnetic
fieldIEC
61000-4-8
Voltage dips, 0 % UT;
Compliance
Electromagnetic environment -
level
guidance
Floors should be wood, concrete
±8 kV contact
or ceramic tile. If floors are
±2 kV, ±4 kV,
covered with synthetic material,
±8 kV, ±15 kV
the relative humidity should be at
air
least 30%.
±2KV for
Mains power quality should be
power supply
that of a typical commercial or
lines
hospital environment.
±1kV Leitung
Mains power quality should be
zu Leitung
that of a typical commercial or
±2kV Leitung
hospital environment.
gegen Erde
Power frequency magnetic fields
should be at levels characteristic
30A/m
of a typical location in a typical
commercial or hospital
environment.
0 % UT;
Mains power quality should be