2.5 Initial calibration

The instrument needs to be calibrated before the first measure-
ment is performed. This is required to set the correct baseline
for the reflectometric optics. For this purpose, a Reflectoquant
Recalibration Set is delivered with the instrument. Please use
EN
the barcode and the plastic test strip for calibration (for details
of the Recalibration Set see Operating Manual). Keep the
instrument at room temperature for at least 30 minutes.
Select [Quality Assurance] in
the [MAIN MENU].
Select [Calibration].
Confirm your selection by pressing
[OK] or [Right (Next)].
Follow the instruction shown on the display.
Correct insertion of the test strip:
Insert the calibration strip of the
recalibration set into the strip adapter
with the beveled edge outwards and
upwards.