DC Voltage
RANGE
400.0 mV
4.000V, 40.00V, 400.0V
600V
NMRR
: >50dB @ 50/60Hz
CMRR
: >120dB @ DC, 50/60Hz, Rs=1kΩ
Input Impedance : 10MΩ, 30pF nominal (1000MΩ for 400.0mV range)
AC Voltage
RANGE
50Hz -- 500Hz
4.000V, 40.00V, 400.0V
600V
CMRR
: >60dB @ DC to 60Hz, Rs=1kΩ
Input Impedance
ACA Current (Clamp-on)
RANGE
50Hz / 60Hz
40.00A
400.0A
600A**
*Accuracy specified to 350A continuous, and 400A for 60 second max with 5 minutes cool down
interval
**Added range to indicate instantaneous over-range current values
Max Hold (Where Applicable)
Specified accuracy ± 50 digits for changes > 25ms in duration
Audible Continuity Tester
Audible threshold : between 20ohm and 120ohm.
Ohms
RANGE
400.0Ω
4.000kΩ, 40.00kΩ, 400.0kΩ
4.000MΩ
40.00MΩ
Open Circuit Voltage : 0.4VDC typical
Capacitance
RANGE*
500.0nF, 5.000µF, 50.00µF, 500.0µF, 3000µF
*Additional 50.00nF range accuracy is not specified
**Accuracies with film capacitor or better
***Specified with battery voltage above 2.8V (approximately half full battery). Accuracy
decreases gradually to 12% at low battery warning voltage of approximately 2.4V
All manuals and user guides at all-guides.com
Elma 3036s + 3038s +
BM126s
: 10MΩ, 30pF nominal
Page 38
Accuracy
0.3% + 4d
0.5% + 3d
1.0% + 4d
1)
Accuracy
1.5% + 5d
2.0% + 5d
1)
Accuracy
1.9% + 8d
1.9% + 8d*
unspecified
Accuracy
0.8% + 6d
0.6% + 4d
1.0% + 4d
2.0% + 4d
Accuracy**
3.5%*** + 6d