measurement errors. Enclosing more than one conductor of a circuit will result in differential
current (like identifying leakage current) measurement.
Strong Electro-magnetic field environments such as high-current transformers, motors and
conductors may affect measurement accuracy.
Temperature Function (BM126s only)
Be sure to insert the banana plug type-K temperature bead probe Bkp60 with correct
polarities. Slide-switch on defaults at degree C (Celsius). Press SELECT button momentarily to
select degree F (Fahrenheit). You can also use a plug adapter Bkb32 (Optional purchase) with
banana pins to type-K socket to adapt other type-K standard mini plug temperature probes.
µ µ µ µ A Current Function (BM125 only)
Inputs are made through the test leads terminals. Slide-switch on defaults at DC. Press
SELECT button momentarily to select AC.
Application notes:
The DC µA function is designed especially for HVAC/R flame sensor applications. The
1)
0.1µA resolution is useful for identifying the minute current changes in flame detector
applications. Flame signal current check should indicate steady flame signal of at least
2µA for a rectification type, or 1.5µA for an ultraviolet type (8µA for self checking systems).