Calibrationprocedure
Whenthe
Action/Description
displayshows
Put the instrument in setup mode
While the software version number is flashing,
1. 1 0
remove the security link and press ENT
Display indicates that unit is /AR type, press ENT
AnLg
Press Å or ì to Calibrate the inputs or ENT to skip
CIPn
Calibrate inputs?
CIPy
Connect the RTD Simulator to the inputs (for connections see over page).
Leave inputs 2 & 3 open circuit
OPEN
Connect 50 Ω resistor
20. 0 0
Connect 200 Ω resistor
5. 0 0
Connect 1 kΩ resistor
1. 0 0
Connect 5 kΩ resistor
0. 2 0
Calibrate RTD Sensor?
rCLy
Set the RTD Simulator to 0 °C
0C
Set the RTD Simulator to 200 °C
200C
Press Å or ì to Calibrate the outputs or ENT to skip to the next setup item
COPn
Whenthe
Action/Description
displayshows
Put the instrument in setup mode and scroll through the main menu
COPn
Press
Calibrate Outputs?
COPy
Measure the output current
The output current should be 0.00 mA
I=0
(Use the keypad to adjust the value as required)
The output current should be 20.00 mA
I=20
Measure the output voltage
The output voltage should be 0.00 volts
E=0
The output voltage should be 10.00 volts
E=10
Instrument will continue with the setup process
rng=
Note: Do not remove the power while the save message is on display.
22
Å
ì
or
ENT
Start
Calibrate
Calibrate
Calibrate
Calibrate
Calibrate
Start
Calibrate
Calibrate
ì
Å
Toggle
Increase Decrease
Increase Decrease
Increase Decrease
Increase Decrease
ï
No
Skip
Skip
Skip
Skip
Skip
No
Skip
Skip
ENT
Start
Accept
Accept
Accept
Accept