USER MANUAL
This note must be read in full. Any operations on live conductors can be dangerous.
•
The operator is expected to be fully aware of all necessary electrical safety regulations and
•
procedures. Safe operation is this responsibility.
It is up to the user to ensure that the equipment is at all times in its original safe conditions.
•
Operate on ranges selector on the bottom part of clamp, free of instrument connection
•
cable, (see Figure) to modify current ranges.
TECHNICAL SPECIFICATIONS
Primary rates current:
Measuring range:
Output signal:
Protection:
Ratio:
Accuracy:
Max error:
Internal load:
External min load:
Frequency range:
Pollution degree:
Dielectric strength:
Temperature range:
Measurement category:
"Opening" cable:
"Opening" bus bars:
Dimensions (mm):
Weight:
Power measurement:
Output connection:
ACCURACY
1A FS
Accuracy in % of the output signal
Phase difference [°]
100A FS
Accuracy in % of the output signal
Phase difference [°]
1000A FS
Accuracy in % of the output signal
Phase difference [°]
Measurements performed with digital instrument, model 1.200.000 digits class 0.2 COFRAC
CAUTION: this symbol indicates that equipment and its accessories shall be subject to a
separate collection and correct disposal.
ENGLISH
CAUTION
1A, 100A, 1000A AC FS
0.001÷1.2A, 0.1÷120A, 1A ÷ 1200A
1V
By diodes
1mV/mA – 10mV/A - 1mV/A
see tables (range 40Hz÷100Hz)
not declared for 101Hz ÷10kHz
2%reading
1A=3kΩ; 100A=15Ω; 1000A=1.5Ω
100 times the internal load
40Hz ÷ 10kHz
2 according to IEC 1010-1
5,5KV 50Hz 1min
-10°C ÷ 50°C (14°F ÷122°F)
CAT III 600V
Diameter 54mm (2in)
35x35mm (1x1in) or 50x12mm (2x1in)
105 x 225 x 31mm (4x9x1in)
720g (25ounces)
Take care to keep the conventional direction P1 P2 rear of the arrows
Shielded cable FM2R 2m long , ended with a FRB D01 connector
1 ÷ 100mA
±(3%rdg+0.5mV)
100mA ÷ 1A
±(1%rdg+1mV)
1 ÷ 10A
±(1%rdg+0.5mV)
traceable.
Ranges selector 1-100-1000A
100 ÷ 500mA
±(2%rdg+1mV)
1 ÷ 10A
±(0.5%rdg+0.5mV)
2
10 ÷ 100A
±(0.5%rdg+1mV)
1.5
HT96U
Versión 2.00 - 06/09/2024
500mA ÷ 1.2A
±(1%rdg+1mV)
<10
10 ÷ 120A
±(0.5%rdg)
1
100 ÷ 1200A
±0,5%rdg
1
2