1. Press SET.
2. Select [CHANNELS] and press CAL.
3. Select [INPUT] and press CAL.
4. Select the desired range and press CAL.
5. Select [ATC] and press CAL.
6. Select the desired temperature compensation input and press CAL.
7. Select [CONTROL] and press CAL.
8. Select the control type (on/off or proportional).
9. Pre-set the desired low level (Lo).
10. Select the corresponding relay (REL).
11. Pre-set the desired high level (Hi).
12. Select the corresponding relay (REL).
13. Pre-set the desired hysteresis (Hy) between the on and off state of the
relays.
14. Pre-set the desired delay (dt) after which the relays should be activated, when
one of the pre-set levels is exceeded.
Proportional control only:
15. Pre-set the desired the proportional control area (dP), in which proportional
regulation should start.
16. Pre-set the desired pulse-time (Pt) of the relays. The relays will pulse at a rate
proportional to the regulation difference.
Optional alarm procedure:
17. Select [ALARM AFTER] and press CAL.
18. The display shows the alarm time after which the process control should be
interrupted in case of a failure. Pre-set the desired alarm-time and press
CAL.
19. Select [REL] and press CAL.
20. Select the desired relay to be closed in case of an alarm and press CAL.
1. Press SET.
2. Select [DATALOGGER] and press CAL.
3. Select [INTERVAL] and press CAL.
4. Select the desired time interval between the data-logging and press CAL.
View the stored values:
1. Select [TABLE], [GRAPH] or [SEND] and press CAL.
2. Select the desired form to display the stored data or send them to a computer
and press CAL to continue. The scales of a graph can be modified by pressing
CAL. Follow the instructions on the screen.
Erase the stored values:
1. Select [ERASE] and press CAL. Follow the instructions on the screen.
1. Press SET.
2. Select [RS485] and press CAL.
3. Select [BAUDRATE] and press CAL.
4. Select the desired baudrate and press CAL.
5. Select [IDENTIF.NUMBER] and press CAL.
6. Select the desired identification number (0...999) and press CAL. The instru-
ment can be identified by e.g. a computer when a specific number is allocated
to it.