Herunterladen Diese Seite drucken

Carlo De Giorgi Sterilintraplant Plus 557/00 Gebrauchs- Und Wartungsanleitung Seite 22

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 46
With reference to Fig. 2, connect the footswitch to socket (4) on the rear
panel by inserting the plug and twisting the locking ring;
Insert the saline solution stand into upper socket (8) and screw it down,
securing it by tightening
required position; now fit the saline solution bottle inside in the plastic
cradle.
6
With reference to Fig. 3 open the side door of peristaltic pump (1) and
remove the tube (keeping it safe for future uses with irrigators other than
the one supplied); insert the larger section tube (2) (included with the
disposable irrigator supplied) into the peristaltic pump, paying attention to:
-
the positioning of supports (3), since, as can be observed in the figure,
the supports perform a restraining function on the exterior of the
pump;
the connection of the "IN" and "OUT" sides (4) respectively to the
-
saline solution bottle and the surgical handpiece (not supplied).
4
GB - 6
the nut (using the supplied wrench) in the
5
3
STERILINTRAPLANT PLUS
7
4
Fig. 2
Fig. 3
2
8
1
Carlo de Giorgi

Werbung

loading

Verwandte Produkte für Carlo De Giorgi Sterilintraplant Plus 557/00

Diese Anleitung auch für:

Sterilintraplant plus 558/00