Herunterladen Diese Seite drucken

Acculab SVI Serie Bedienungshandbuch Seite 7

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

9
To exit "Counting" press and hold "TARE/ZERO" key for more than 3 seconds to access application
menu and select "UNIT" (default setting).
Totaling:
1
To select the application "Totaling" press and hold "TARE/ZERO" key for more than 3 seconds to access
application menu.
2
With the "F" key you can scroll through the menu options (UNIT, COUNT, & TOTAL).
3
Select "TOTAL" by pressing the "CAL/PRINT" Key.
4
Display will automatically return to weighing mode with "TTL" displayed (signifying to the end user that
they are in application mode).
5
Place 1st object on scale and press "F" (value is stored) / remove object and display will return to zero.
Place 2nd object on scale and press "F" (value is added to 1st) and the "TTL" symbol will flash (indicating
6
total weight stored) / remove object and display will return to zero ("TTL" stops flashing).
7
Place 3rd object on scale and press "F" (value will be added and "TTL" will flash) / remove object and
display will return to zero ("TTL" stops flashing) process can be repeated up to 99 times.
8
Holding down the "F" key for 3 seconds will clear all memory.
9
To exit "Totaling", press and hold "TARE/ZERO" Key for more than 3 seconds to access application menu
and select "Unit" (default setting).
Features:
Digital Calibration:
Allows for convenient user calibration from the keypad with three selectable weight options.
Power Up Test:
When the unit is initially turned on all segments of the display will appear. The segments will be displayed for
approx. 3 seconds. After self-test scale will return to zero.
Overload:
When a load or force is applied to the scale platform that exceeds the unit's capacity "H" will appear in the
d isplay indicating an overload has occurred. Immediately removing the object or weight from the unit will allow
normal operation to return. If unit does not return to normal operation, damage has occurred.
Auto Off:
During battery operation (in order to conserve battery consumption) the scale will automatically turn off after
approximately 4 minutes of inactivity / default setting. (Please see "Menu" options for additional settings).
RS-232:
Standard feature included with scale. (Please see Interface Description)
Care and Maintenance:
Avoid lengthy exposure to extreme heat or cold. Please allow the scale to acclimate to room temperature
for at least 15 to 20 minutes if exposed to extreme temperatures. (Please see Product Specifications)
Keep product clean. Remove pan from base and clean accordingly. Do not "wash down" scale or indicator
with water or cleaning liquids.
Please use supplied packaging when transporting scale.
Remember to remove battery when not in use.
ENGLISH
4

Werbung

loading

Diese Anleitung auch für:

Svi-10aSvi-20bSvi-50c