Herunterladen Diese Seite drucken

neno Sano Bedienungsanleitung Seite 25

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
Voltage
0% UT; 0.5
cycle
drops, short
at O°, 45°,
interruptions
90°, 135°,
and voltage
180°, 225°,
variations on
270° i
supply lines
315°
IEC 610004-11
3 Vrms
Conducted RF
150 kHz to
current IEC
80 MHz
61000-4-6
10VAn
Radiated radio
frequency-
80 MHz to
EC6100CM-3
2.7 GHz
0% UT; 0.5 cycle
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
3 Vrms
10 V/m
The quality of the mains
power supply should be
that of a typical commercial
or hospital environment.
If the user of the NENO
SANO Compressor Nebuliser
(VP-D2) requires continuous
operation during power
outages, it is recommended
that the NENO SANO Com-
pressor Nebuliser (VP-D2 ) is
powered by an emergency
power supply or battery.
The manufacturer should
consider reducing the
minimum separation
distance, based on RISK
MANAGEMENT, and using
higher LEVELS OF RE-
SISTANCE TESTING that are
appropriate for the reduced
minimum separation distan-
ce. Minimum separation
distances for higher LEVELS
OF RESPONSE TESTING
should be calculated using
the following equation:
E= 6/d √ P
Where P is the maximum
power in W. „d" is the mini-
mum separation distance in
m and E is the immunity test
level in V/m.
25