e. Connect one end of the humidifier tube to the lid of the
humidifier bottle.
f. Connect the other end of the humidifier tube to the outlet barb
of the CAIRE Companion 5.
g. Place the bottle in its designated space on the Humidifier
Bottle Support Stand.
h. Secure the bottle on the stand using the elastic band.
i. Proceed to step 4.
Step 4: Power On and Warm Up
a. Plug the electrical cord of the CAIRE Companion 5 into an
AC outlet.
WARNING: INSPECT THE ELECTRICAL CORD FOR DAMAGE
BEFORE USE. IF THE CORD IS DAMAGED, DO NOT PLUG IT
INTO AN ELECTRICAL OUTLET OR ATTEMPT TO OPERATE
THE CONCENTRATOR.
b.Press the power switch in the "ON" (I) position. When the
Companion 5 is powered on properly, the green indicator on
the LED Display will light up. All LED lights will illuminate
upon start-up. After the concentrator completes the warm-up
cycle, only the green light will remain on.
Note: All LED lights will illuminate upon start-up. After the concentrator
completes the warm-up cycle, only the green light will remain on.
For OCSI Models: The O2 Light will illuminate for four (4)
seconds upon start-up. After this initial start-up, this light
is disabled for approximately five (5) minutes while the
concentration rises to specifications.