Quick Start Guide
Charge Amplifier and Data Acquisition Device
Types 5165A... & 5167A...
A
A
B
A
B
C
Foreword
Thank you for choosing a Kistler quality product. Please read these instructions
carefully, so that you can take optimum advantage of the versatile features of
this product.
The information in this document is subject to change at any time without
prior notice. Kistler reserves the right to improve and modify the product in
accordance with technical progress without the obligation to inform persons
and organizations based on these changes.
©2019 Kistler Group. All rights reserved.
Contents
1. General Information
2. Device Description
3. Connection Setup via Ethernet
4. Control via Web Interface, Instruction Manual
5. Activation of purchased software features
6. Restoring all Settings
7. Repairs at Kistler
8. Disposal Instructions for Electronic Devices
1.
General Information
The Kistler LabAmp Type 5165A... and 5167A... are universal laboratory
measuring amplifiers. Signals from piezoelectric sensors – and in case of the
5165A also from Piezotron® sensors (IEPE) and voltage signals – are digitized
and processed with high resolution. Data acquisition functions as well as
freely configurable analog outputs emphasize the comprehensive application
options of the device.
The laboratory measuring amplifier is configured as well as operated with an
intuitive web interface that is easily available through an external computer
via a standard web browser (Minimal requirements: Chrome 43, Firefox 38,
IE 11). The web browser-based data download permits easy further processing
of the acquired signals in an analysis software.
The present instructions are for rapid commissioning of the device.
For initial commissioning of the device, see chapter 3.
2.
Device Description
The following brief description explains the basic components of the measur-
ing amplifier and refers to the corresponding chapters with more detailed
descriptions:
Front side – 5165A... & 5167Ax0
E
E
K
C
D
Front side – 5167Ax1
E
C
K
D
Back side – 5165A... & 5167A...
B
J
A
F
G
Kistler Group
Eulachstrasse 22
8408 Winterthur, Switzerland
Description
Connector of the 24 V power supply
A
B
(e.g. plug-in power supply unit)
A
B
C
Main switch for the power supply of the device
Sensor inputs:
A
B
C
D
• BNC neg. → Type 5165A... & 5167Ax0
• Fischer 9-pole neg. → Type 5167Ax1
E
A
B
C
D
Analog outputs (BNC neg.)
Light-emitting diodes (LEDs), indicating the status
E
A
B
C
D
F
of the corresponding sensor input or analog output
Remote control connector (D-Sub 9-pole neg.) →
Type 5167A... only:
• Measure/Reset control (Pin 5)
E
A
B
C
D
F
G
• Start/Stop trigger (Pin 4)
• 24 V supply (Pin 1)
• GND (Pins 2 & 6)
USB ports (not supported):
E
B
C
D
F
G
H
• 1x USB → Type 5165A...
• 2x USB → Type 5167A...
2 Ethernet interfaces (RJ45) with switch functionality
E
C
D
F
G
H
I
for connection to a network or computer
516xA_002-800m-09.19
Reset button for resetting the device to
E
D
F
G
H
I
factory settings
J
K
Ground lug
J
K
Fold-out feet
2.1 Status LEDs
The LEDs indicate the status of the corresponding sensor input or analog out-
put and the general operating statuses of the device.
The table below shows the possible statuses:
LED Status
3.
Connection Setup via Ethernet
The measuring amplifiers Type 5165A... and 5167A... include an integrated
web server, whose web interface is used to operate the device. Please ensure
that your PC is in DHCP mode and/or Auto-IP mode (network settings).
The following approach is recommended to establish a connection with the
measuring amplifier via Ethernet:
1. Connect the Device to the Network
Connect the measuring amplifier (connector H –> Sync In or Sync Out) via
a network cable directly to a computer or a network (e.g. network switch).
2. Find the Device in the Network
a. Start the „Network Setup Wizard" tool that finds all Kistler network
devices (tool can be found under www.kistler.com in the download area
of the Kistler LabAmp Type 5165A or 5167A).
b. Click on „Network Setup Wizard" and wait until Kistler network devices
are displayed. In the case of a direct connection, it may take up to
2 minutes until the devices are displayed.
c. The table displays now all identified Kistler network devices.
3. Access to the Device via Web Browser
K
Choose your device in the table by clicking on the table entry and select one
of the following two options:
a. "Open Browser"
Opens the web interface of the respective device in a standard web
browser. The measuring amplifier can now be configured and operated
via the web interface (–> Chap. 4).
E
b. "Identify"
Permits an identification of the respective device. The identified mea-
suring amplifier signals this through the yellow flashing of all LEDs for
approx. 4 s (–> Chap. 2.1). This function is very useful if several mea-
suring amplifiers are used in the same network.
K
Adapting Network Settings:
The network settings for the marked device can be adapted by clicking on
„Next". The following options can be configured for this purpose:
• Device name:
H
I
The device name is displayed in the "Network Setup Wizard" as well as in
Tel. +41 52 224 11 11
Fax +41 52 224 14 14
info@kistler.com
Chapter
E
F
G
H
C
D
I
E
D
F
G
H
I
E
F
G
H
I
F
G
H
I
G
H
I
2.1
H
I
I
6
A
B
C
D
Operating Status
Blue
Sensor input is measuring /
illuminated
analog output is active
Sensor input in reset state
Blue flashing
(Type 5167A... only)
Red
Error (e.g. range overflow, open
illuminated
IEPE input, etc.)
Analog output is switched off,
Switched off
0V at the output
LEDs are illumi-
Initialization after switch-on of
nated in yellow
the device or software update
from left to right
Temporary status (e.g. device
Yellow flashing
identification, read TEDS, etc.)
www.kistler.com
the web interface. In networks that support name resolution, the device
name can be entered directly in the address field of the web browser to
access the device.
• Address mode:
The address modes "DHCP with Auto-IP" and „Static IP" are supported.
It is necessary to click „Next" and then "Update" to ensure that the network
settings are accepted.
Information Regarding Network Connection:
The "Network Setup Wizard" finds Kistler network devices
throughout the network. However, if there are several routers
or switches between the computer and the measuring amplifier,
this search might not be successful under certain circumstances.
4.
Control via Web Interface, Instruction
Manual
The measuring amplifier is operated via the web interface. After the first con-
nection with the device (–> Chap. 3), the home screen is displayed.
The user manual which describes all the functionality of the web interface is
accessible by clicking on the "?" in the upper right corner of the web interface.
E
F
G
H
I
5.
Activation of purchased software features
(from FW version 1.3.0 on)
To activate purchased software features, it is recommended to connect the
device to a computer or network with Internet access (online activation). If the
network installation does not allow this, an offline activation (device without
access to the Internet) is also possible.
Online Activation
For the online activation of purchased software features, proceed as follows:
1. Establish connection to the device via Ethernet, if not already done
( –> Chap. 3)
2. Click on „Activate License" in the LabAmp web interface under menu
„Maintenance" – „Software Licensing"
3. Confirm note „License Activation" with „Next"
4. Note „License activated" appears and can be confirmed with „Close"
Offline Activation
To activate purchased software features offline, proceed as follows:
1. Establish connection to the device via Ethernet, if not already done
( –> Chap. 3)
2. Click on „Activate License" in the LabAmp web interface under menu
„Maintenance" – „Software Licensing"
3. Confirm note „License Activation" with „Next"
4. Follow the instructions until the activation is complete
6.
Restoring all Settings
The device can be reset to the factory settings by using the Reset button (I).
This causes all user and network settings to be reset.
To reset the settings, the Reset button (I) on the back of the device must be
actuated with a pointed object (e.g. needle) for at least 2 seconds.
7.
Repairs at Kistler
Repairs at the Kistler factory can be arranged via the local sales company.
Information can be found at www.kistler.com
8.
Disposal Instructions for Electronic Devices
Old electronics devices may not be disposed of
with household refuse/residual waste. Please return
obsolete devices to the nearest electronics dis-
posal center for disposal or contact your Kistler Sales
Representative.

Werbung

loading

Inhaltszusammenfassung für Kistler 5165A Serie

  • Seite 1 The information in this document is subject to change at any time without Fold-out feet – prior notice. Kistler reserves the right to improve and modify the product in accordance with technical progress without the obligation to inform persons and organizations based on these changes.
  • Seite 2 Eigenschaften Ihres Produkts optimal nutzen können. Masse-Anschluss – Die Angaben in diesem Dokument können jederzeit ohne Vorankündigung geändert werden. Kistler behält sich das Recht vor, das Produkt im Sinne des Ausklappbare Standfüsse – technischen Fortschritts zu verbessern und zu ändern, ohne Verpflichtung, Personen und Organisationen aufgrund solcher Änderungen zu benachrichtigen.