Herunterladen Inhalt Inhalt Diese Seite drucken

Error Codes - A&D GX-L-Serie Anleitung

Hochleistungs-präzisionswaage
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 26
EN.
5.

Error Codes

Display
Error code
Internal error
If this error continues to be displayed, please request repair from your
local dealer.
Stability error
The balance is unable to stabilize due to the environment. Check
around the pan. Prevent vibration, drafts, temperature changes, static
EC, E11
electricity and magnetic fields from affecting the balance. Refer to "2-
3. During Use" in the separate manual. To return to weighing mode,
press the CAL key.
Outside of the setting range
The data to be stored is outside of the setting range. Set again within
the setting range.
Balance internal memory element malfunction
If this error continues to be displayed, please request repair from your
local dealer.
Internal mass error (GX-L series only)
Applying the internal mass does not yield a change in the mass value
EC, E16
as specified. Confirm that there is nothing on the pan and perform the
operation from the beginning again.
Internal mass error (GX-L series only)
The internal mass application mechanism does not function properly.
EC, E17
Perform the operation from the beginning again. If the problem
persists, please request repair from your local dealer.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, please ask for repair to your
local dealer.
Balance internal memory abnormality
If this error continues to be displayed, please request repair from your
local dealer.
Sensitivity adjustment weight error (Positive value)
The sensitivity adjustment weight is too heavy. Check around the
EC, E20
pan. Check the mass value of the weight.
To return to the weighing mode, press the CAL key.
Sensitivity adjustment weight error (Negative value)
The sensitivity adjustment weight is too light. Check around the pan.
EC, E21
Check the mass value of the weight.
To return to the weighing mode, press the CAL key.
Overload error
A sample beyond the balance weighing capacity has been placed on
the pan. Remove the sample from the pan. If the problem persists,
please request repair from your local dealer.
Weighing pan error
The mass value is too light. Check that the weighing pan is installed
correctly. Set the weighing pan correctly. Adjust the sensitivity of the
balance.
Overload error
A sample beyond the balance weighing capacity has been placed on
the pan. Remove the sample from the pan. If the problem persists,
please request repair from your local dealer.
Weighing pan error
The mass value is too light. Check that the weighing pan is installed
correctly. Set the weighing pan correctly. Adjust the sensitivity of the
balance.
Description