WARNING
Using this unit for medical
purposes
The connectors on this unit are not isolated.
Do not connect any device other than one
which conforms to IEC 60601-1 standards.
When an information technology device or
AV device that uses an alternating current is
connected, current leakage may result in an
electric shock to the patient or operator.
If use of such a device is unavoidable,
isolate its power supply by connecting an
isolation transformer, or by connecting an
isolator between the connecting cables.
After implementing these measures,
confirm that the reduced risk now conforms
to IEC 60601-1 standards.
IMPORTANT
The nameplate is located on the bottom.
Caution
Do not use the device in a MR (Magnetic
Resonance) environment.
It may cause a malfunction, fire, and
unwanted movement.
Caution
A paper cutter is mounted
with the unit. When loading
paper, be careful not to touch
the paper cutter. Touching the
paper cutter may cause injury.
10
GB
Caution
Do not insert your finger into
the back of the paper tray.
Your finger may become
stuck in an internal gap,
which may lead to injury.
Caution
Do not touch any unit circuitry and the
patient at the same time.
If the unit malfunctions, it may generate
voltage that could be harmful to the patient.
Caution
• To prevent internal heat built-up,
maintain sufficient space around the
printer (at least 10 cm (4 in.)). Do not
obstruct the ventilation holes on the right
and left sides of the printer and on the
rear panel.
• If the space around the unit is less 10 cm
(4 inches), when used in conjunction
with other devices or for other reason,
keep the vents on the left, right, and rear
sides of the printer clear as much as
possible and ensure that the ambient
temperature is maintained below 40 °C.
If necessary, install a cooling fan to
ensure ventilation of the interior of the
unit to prevent the risk of heat
confinement before using the printer.
Warning on power
supply devices
For the DC power supply, be sure to use a
device that complies with the IEC 60601-1
standard and is protected by double
insulation or reinforced insulation.
Be sure to use a power supply device
capable of supplying a voltage of 12 V to
24 V and a current of 8.2 A (at 12 V) to
3.7 A (at 24 V) or more.
However, the device should not be capable
of supplying 240 VA for 60 seconds or
more (20 A or less at 12 V, 10 A or less at
24 V).